22
Eco/Classic Installation Guide and Technical Specification
3.2
Handset with Locking Function
The motorised bed functions can be operated via the handset. All functions can be locked
with the nurses’ key.
To avoid damage, the handset should always be hung up on the handset hook (e.g.
mattress base or side rail) when not in use.
ATTENTION:
Press only one button at a time, as the system could overload and become damaged.
3.3
Locking Function for Handset
On the back of the handset there are two locking devices, which are operated using the
supplied nurses’ key. The upper lock restricts the Trendelenburg feature, whilst the lower
lock restricts all electric adjustment functions.
The switching positions I and II are testing settings, used to check the safety during the
annual inspection, after repair work, or each time the bed is put into service again.
Backrest adjustment
up/down
Handset hook
Thigh rest adjustment
up/down
Mattress base height
up/down adjustment
Trendelenburg
Trendelenburg Lock
Nurses’ key
Locking device