- 25 -
Flashing
Low charge remaining.
This indicates the remaining battery charge for a few measurements.
Remaining
No charge remaining.
The device must be charged for measurements to be possible.
Inflation of the cuff started again during deflation (during
measurement)
When, for example, the set inflation value is insufficient for blood pressure measurement, the patient
moves during measurement, or noise is detected, inflation occurs again until the cuff reaches a value
higher than the preset pressure value after deflation starts.
This does not indicate a problem.
Automatic re-inflation: See page 3
Make sure that the patient stays still during measurement.
Blood pressure is different each time or extremely high (low)
The measurement posture is different or the measurement position does not match the height of the heart.
Make sure the patient is in the correct posture for measurement.
Make sure that the measurement position matches the height of the heart.
Blood pressure values differ by psychosomatic states.
Measure under the same conditions.
The battery mark is displayed when a full LCD display occurs when the power switched on. The
display at this time does not indicate that the remaining charge is insufficient.