- 14 -
If the preset pressure value was insufficient for
the measurement, inflation is carried out again
automatically.
For information about re-inflation, see page 3.
5. Connect the cuff
Measure the patient’s arm circumference, and then use
an appropriate cuff based on the table on the right.
Securely insert the air plug of the cuff into the air
connector on the left side of the main unit.
6. Apply the cuff to the patient’s arm.
Find a chair and a table so that the patient can be
comfortably seated with patient’s back and arm
supported. Legs should not be crossed and feet kept flat
on the floor.
Measure the left arm during AUTO measurement. If
the right arm is measured, it may not be possible to
measure normally due to the attachment state of the cuff.
When applying the cuff, make sure that the ARTERY
mark “
” aligns with the brachial artery and the
INDEX mark “
” is within the OK line.
7. Press the START/STOP BUTTON to start
measurement
Measurement
count
CUFF SIZE
REGULAR
LARGE
ARM CIRCUMFERENCE
22-32 cm
32-42 cm
2 to 3cm
AIR PLUG