- 17 -
5. Press the SET/MEMORY BUTTON to set the
deflation rate
The deflation rate displayed in the bottom left of the
display area changes each time the SET/MEMORY
BUTTON is pressed.
Select from one of the following: 2.5, 3.5, 4.5, 5.5, or 6.5
mmHg/s.
6. Connect the cuff
Measure the patient’s arm circumference, and then use an
appropriate cuff based on the table on the right.
Securely insert the air plug of the cuff into the air
connector on the left side of the main unit.
7. Apply the cuff to the patient’s arm.
Find a chair and a table so that the patient can be
comfortably seated with patient’s back and arm supported.
Legs should not be crossed and feet kept flat on the floor.
As when measuring with a standard column of mercury,
apply the cuff, and then apply the stethoscope.
When applying the cuff, make sure the INDEX mark
“
” is within the OK line.
Deflation rate
CUFF SIZE
REGULAR
LARGE
ARM CIRCUMFERENCE
22-32 cm
32-42 cm
2 to 3cm
AIR PLUG