3.8 Display statistical data 

 

 
 

 

 

(nur bei FMT-W30) 

S

 

>>> 

 

S

 

>>> 

 

After you finish the (last) measurement you can recall 
the statistical results by pressing the S-Button. 

 

 

 

 

S

 

>>>

 

 

 

Highest value of all saved tests. 

 

 

 

S

 

>>>

 

 

 

Lowest value of all saved tests. 

 

 

 

 

S

 

>>>

 

 

 

Average (arithmetic medium) of all saved tests. 

 

 

 

S

 

>>>

 

 

Standard deviation related to the average. 

 

Using the optional Software FMI-S_Connect the statistic data can be replenished with the instruments base data and the work order data. If the 
USB interface is connected the single test values can be transferred and documented also. To capture single test values of the manual pull 
tester FMT-W30, the USB connection to your PC has to be established during the entire test cycle. 

 
 

3.9 Delete displayed data and statistics 

To delete the actually displayed data you press the 0-Button, the display shows the 
value “0.0”. By pressing the 0-Button for a longer time (>2s) you delete all memorized 
data at the same time. Subsequently you can change the setting or load new test 
limits. 

 

The memory and static function can be turned off. (see chap. 6) 

 

 

0

2S 

 

 
 

4. Tensile force limit control 

Depending on the wire strength and the related standard, the following tensile strength should at least be achieved. All below mentioned 
standards can be performed with a pull speed of 100mm/min (4“/min), the default speed oft he motorized pull tester FMT-W40. Utilizing the 
optional software FMT-W_Connect other pull speeds can be setted. 

 

AWG 

Cross section 

Cable diameter 

 

SAE AS7928 Table II 

IEC 60352 Teil 2 

UL 486 C 

NASA Std 8739.4 

30 

0,06 mm² 

0,36 mm 

 

 

6 N 

6 N 

 

28 

0,09 mm² 

0,38 mm 

 

 

11 N 

11 N 

22 N 

26 

0,14 mm² 

0,48 mm 

 

32 N 

18 N 

18 N 

36 N 

24 

0,22 mm² 

0,61 mm 

 

45 N 

28 N 

28 N 

36 N 

22 

0,34 mm² 

0,76 mm 

 

67 N 

40 N 

40 N 

57 N 

20 

0,56 mm² 

0,97 mm 

 

85 N 

60 N 

45 N 

92 N 

18 

0,93 mm² 

1,27 mm 

 

170 N 

90 N 

45 N 

142 N 

16 

1,25 mm² 

1,44 mm 

 

223 N 

135 N 

68 N 

183 N 

14 

1,93 mm² 

1,80 mm 

 

312 N 

200 N 

100 N 

290 N 

12 

3,16 mm² 

2,29 mm 

 

490 N 

275 N 

138 N 

459 N 

10 

4,65 mm² 

3,10 mm 

 

 

355 N 

 

707 N 

 

Remarks: DIN 

41611/3 is replaced by DIN IEC 60352 Part 2 

MIL-T-7928 is replaced by SAE AS7928 Table II 

BS5B178 corresponds to IEC 60352 Part 2 

UL486A corresponds to IEC 60352 Part 2 

Summary of Contents for FMT-W30C5

Page 1: ...re Terminals Dynamom tre Numerique par Jonctions de C bles Dinamom trico Digital Por cables con unions Adecuadas Dinamometri Digitales a di Cavi con Connettori BDA Version 2 4 Bedienungsanleitung Oper...

Page 2: ...l tester FMT W30 20 3 6 Start measuring AutoTara with motorized pull tester FMT W40 21 3 7 Automatic memorizing of measured values with motorized pull tester FMT W40 21 3 8 Display statistical data 22...

Page 3: ...formation 2 1 Scope of Supply unpacking and Setting Up The scope of supply consists of Measuring instrument with integrated electronics Hand lever or roller grip USB Interface cable for optional softw...

Page 4: ...ecommended date of next calibration and the nominal measuring range Fn To turn the instrument off you may press the I Button for 2 seconds else the AUTO OFF function will turn the instrument off autom...

Page 5: ...he cable connector stays securely inside the ring and cannot be pulled out through the slot The wire clamp fixture should be positioned in the full open position The manual pull tester needs a free le...

Page 6: ...e develops which works upon the cable connection The peak value which mostly appears before the connection snaps up is captured at 1000Hz and is indicated in the display After 5 seconds of unchanged m...

Page 7: ...tive if 25 of the force limit minimum tensile strength P3 or a force of 40N is reached If the actual measured force falls below this level the cable clamp returns to its home position and the measurin...

Page 8: ...etting or load new test limits The memory and static function can be turned off see chap 6 0 2S 4 Tensile force limit control Depending on the wire strength and the related standard the following tens...

Page 9: ...not apply with the motorized version FMT W40 The lower limit is adjusted automatically to 25 or max 40N of the upper limit With a short press on the S Button you return to the main setting menu I 0 Pr...

Page 10: ...icense file has to be generated and saved in the Alluris folder Please enter the received license key and the serial no of your pull tester and click generate license file The license key has to match...

Page 11: ...standard requiring 30 N the instrument is parametrized to 36 N This allows pulling further while testing than the limit set to compensate for a relaxing of the test object The test report refers to t...

Page 12: ...mpany details with your own logo etc The template is filed in C Alluris FMT W_Connect Formular xlsx Fields with gray background are filled by the data entered in FMT W_Connect and the data of the test...

Page 13: ...to the main setting menu 0 I 1 0 The pull tester tares and resets the PEAK automatically with start of each measurement in order to eliminate any influences by change of the material and temperature I...

Page 14: ...a PC 24 VDC 3A universal power supply Standby consumption 42mA Average consumption drive active 0 8A Interface USB 2 0 Optional software FMT W_Connect Temperature range Operation 5 40 C 5 30 C Storage...

Page 15: ...29 8 2 Register for updates In order to receive information about product changes and updates and to enjoy full warranty please register your product at www alluris de service php...

Page 16: ...service A loud vibrating sound occurs but the clamping device does not turn and the pull force does not increase They drive might be blocked and overloaded Switch the instrument off and remove the blo...

Page 17: ...irm additionally that the product complies with the requirements of the EMC Directive 2006 42 EEC In bereinstimmung mit der WEEE Richtlinie 2002 96 EC ist dieses Ger t eingestuft als Monitoring and Co...

Page 18: ...nischen Bundesanstalt PTB Braunschweig im Rahmen des DAkkS zur ckf hren A 2 Calibration Confirmation acc DIN EN 10204 2 1 We hereby confirm in accordance to DIN EN 10204 2 1 that this instrument has b...

Page 19: ...75 Alluris GmbH Co KG Basler Strasse 63 65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44...

Page 20: ...65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44 BDA_FMT W Art Nr 50337 Stand 07 15 Alluris GmbH Co KG Alle Rechte vorbehalten Alluris ist eine ei...

Reviews: