28 

 

 

With a long press on the S-Button (>2 seconds) you leave the setting menu and return to the main operating 
menu. 

 

 

 
7. Technical data 

 

FMT-W30C5 FMT-W30K1  FMT-W40C5 

FMT-W40K1 

Measuring Range 

Range in Newton (N) 

0...500 N 

0…1000 N 

0…500 N 

0…1000 N 

Alternative units 

N | kg | lb 

Wire Diameter

 

Slot width terminal adapter 

0,5 | 0,8 | 1,0 | 1,3 | 1,5 | 2,0 | 2,5 | 3,0 | 3,5 | 4,0 | 5,0 | 6,0  mm 

Clamp opening 

0,1 ... 6,0 mm (continously variable) 

Dia. Range  IEC 60352-2 

0,05 ... 4 mm² 

(AWG10...30) 

0,05 ... 6 mm² 

(AWG8...30) 

0,05 ... 4 mm² 

(AWG10...30) 

0,05 ... 6 mm² 

(AWG8...30) 

Dia. Range SAE AS7928 II  

AWG 14 ... 28 

AWG 10 ... 28 

AWG 14 ... 28 

AWG 10 ... 28 

Resolution 

 

0,5N (0,05kg / 0,1lb) 

1 N (0,1kg / 0,2lb) 

0,5 N (0,05kg / 0,1lb) )* 

0,5 N (0,05kg / 0,1lb) 

1 N (0,1kg / 0,2lb) 

0,5 N (0,05kg / 0,1lb) )* 

Accuracy 

@ 23°C (F.S.) 

+/- 0,5% (+/- 1 increment of resolution) 

Tk (absolute) 

selfadjusting at start 

Tk (relativ)e 

+/- 0,02% (°K) 

Betriebsarten 

Standard  

displays real time value in N | kg | lb 

PEAK 

displays peak value in N | kg | lb (selectable) 

Pulls until break 

 

Yes 

Pull test with dwell time 

 

Yes )* 

Pull-Release (until force limit) 

 

Yes )* 

Pull Speed 

Speed 

 

10 | 25 | 50 | 100 | 150 | 200 mm/min )* 

Accuracy 

 

+/- 5% 

Overload 

max. admissible 

200 % (F.S.)  |  alarm at 120% (F.S.) 

Display 

Type 

LCD, 5-digit, 12mm high 

Update (Standard) 

1000 msec | 500 msec | 333 msec | 200 msec |100 msec | 50 msec (selectable) 

Update (Peak) 

ca. 1 msec 

Memory 

Manual memory 

Capacity of 1000 test cycles  | Statistics with display of average, maximum, 

minimum and standard deviation 

Auto memory 

 

Yes )* 

Power supply 

Type 

Solar cell 

With USB connected via PC 

24 VDC | 3A universal power supply 

Standby consumption: 42mA 

Average consumption (drive active):  0,8A 

Interface 

USB 2.0 

Optional software FMT-W_Connect 

Temperature range 

Operation 

5°…40° C 

5°... 30° C 

Storage 

-20°... 60° C (rF < 80%) 

Protection code 

 

IP 40 

Weight 

 

app. 14 kg 

Dimensions 

LxWxH (without lever) 

350x160x75 mm 

320 x 160 x 115 mm 

Housing material 

 

Main plate surface hardened steel, Grip and 

fixtures stainless steel, cover anodized aluminum 

Main plate surface hardened steel, Grip and fixtures 

stainless steel, cover anodized aluminum and 

polystyrene. 

)* function or setting requires software FMT-W_Connect  

8. Maintenance and calibration 

 
 
This instrument is maintenance-free under proper use. Torn parts of wire must not fall into the spaces 
around the connection adapter; should this occur, blow them off or remove them using the vacuum cleaner. 
 
Force gauges should be inspected and re-calibrated on a regular base, at least once a year. Our calibration 
services comprise the inspection, and calibration according ISO 9001:2008 approved procedures and a 
readjusting of the load cell if necessary. (E-mail address: 

[email protected]

). 

 
 
 
 
 

8.1 5-Years extended warranty 

We grant for all Alluris force gauges a 5 years extended warranty period starting with the date of purchase, subject to the registration of the 
device before it is taken into operation and the service and calibration at regular intervals in our calibration laboratory. Consumption material, 
normal wear and tear as well as damages caused by improper use are excluded from this warranty. Apart from that the warranty clause of our 
general conditions of sales and delivery are valid. 

 

Summary of Contents for FMT-W30C5

Page 1: ...re Terminals Dynamom tre Numerique par Jonctions de C bles Dinamom trico Digital Por cables con unions Adecuadas Dinamometri Digitales a di Cavi con Connettori BDA Version 2 4 Bedienungsanleitung Oper...

Page 2: ...l tester FMT W30 20 3 6 Start measuring AutoTara with motorized pull tester FMT W40 21 3 7 Automatic memorizing of measured values with motorized pull tester FMT W40 21 3 8 Display statistical data 22...

Page 3: ...formation 2 1 Scope of Supply unpacking and Setting Up The scope of supply consists of Measuring instrument with integrated electronics Hand lever or roller grip USB Interface cable for optional softw...

Page 4: ...ecommended date of next calibration and the nominal measuring range Fn To turn the instrument off you may press the I Button for 2 seconds else the AUTO OFF function will turn the instrument off autom...

Page 5: ...he cable connector stays securely inside the ring and cannot be pulled out through the slot The wire clamp fixture should be positioned in the full open position The manual pull tester needs a free le...

Page 6: ...e develops which works upon the cable connection The peak value which mostly appears before the connection snaps up is captured at 1000Hz and is indicated in the display After 5 seconds of unchanged m...

Page 7: ...tive if 25 of the force limit minimum tensile strength P3 or a force of 40N is reached If the actual measured force falls below this level the cable clamp returns to its home position and the measurin...

Page 8: ...etting or load new test limits The memory and static function can be turned off see chap 6 0 2S 4 Tensile force limit control Depending on the wire strength and the related standard the following tens...

Page 9: ...not apply with the motorized version FMT W40 The lower limit is adjusted automatically to 25 or max 40N of the upper limit With a short press on the S Button you return to the main setting menu I 0 Pr...

Page 10: ...icense file has to be generated and saved in the Alluris folder Please enter the received license key and the serial no of your pull tester and click generate license file The license key has to match...

Page 11: ...standard requiring 30 N the instrument is parametrized to 36 N This allows pulling further while testing than the limit set to compensate for a relaxing of the test object The test report refers to t...

Page 12: ...mpany details with your own logo etc The template is filed in C Alluris FMT W_Connect Formular xlsx Fields with gray background are filled by the data entered in FMT W_Connect and the data of the test...

Page 13: ...to the main setting menu 0 I 1 0 The pull tester tares and resets the PEAK automatically with start of each measurement in order to eliminate any influences by change of the material and temperature I...

Page 14: ...a PC 24 VDC 3A universal power supply Standby consumption 42mA Average consumption drive active 0 8A Interface USB 2 0 Optional software FMT W_Connect Temperature range Operation 5 40 C 5 30 C Storage...

Page 15: ...29 8 2 Register for updates In order to receive information about product changes and updates and to enjoy full warranty please register your product at www alluris de service php...

Page 16: ...service A loud vibrating sound occurs but the clamping device does not turn and the pull force does not increase They drive might be blocked and overloaded Switch the instrument off and remove the blo...

Page 17: ...irm additionally that the product complies with the requirements of the EMC Directive 2006 42 EEC In bereinstimmung mit der WEEE Richtlinie 2002 96 EC ist dieses Ger t eingestuft als Monitoring and Co...

Page 18: ...nischen Bundesanstalt PTB Braunschweig im Rahmen des DAkkS zur ckf hren A 2 Calibration Confirmation acc DIN EN 10204 2 1 We hereby confirm in accordance to DIN EN 10204 2 1 that this instrument has b...

Page 19: ...75 Alluris GmbH Co KG Basler Strasse 63 65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44...

Page 20: ...65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44 BDA_FMT W Art Nr 50337 Stand 07 15 Alluris GmbH Co KG Alle Rechte vorbehalten Alluris ist eine ei...

Reviews: