background image

27 

6. General settings (P1), Setting of memory (P2) and reset works defaults (Po) 

To customize the instrument to the application, some of the factory settings (defaults) can be changed. These adjustments will be saved in the 
microprocessor and be recalled when you switch on the force gauge the next time. In order to change the setting you press the S-Button for app. 
2 second until the code P1 is displayed. Press the I-Button to choose the submenu. 
 

 

 

 

kg 

lb 

In menu P11 you can select the measuring unit by pressing the I-
Button. The chosen unit flashes in the display. Default setting is the 
SI-dimension Newton (N). Pressing the 0-Button you change to the 
next submenu. 

 

 

 

 

With a short press on the S-Button you return to the main setting menu. 

 

 

 

 



10 

The load cell captured the force with a rate of 3,6kHz. The display 
update is limited to 10Hz in order to make the display readable. You 
may reduce this rate by selecting an appropriate value with the I-
Button.  Pressing the 0-Button you change to the next submenu. 

 

 

 

 

With a short press on the S-Button you return to the main setting menu. 

 

 

 

 

10 

20 
30 

The measurement will be stopped automatically if no new force 
values are captured for 5 seconds. You may change this period by 
pressing the I-Button. Pressing the 0-Button you change to the next 
submenu. (Applies for FMT-W30 only) 

 

 

 

 

With a short press on the S-Button you return to the main setting menu. 

 

 

 

 

1|2 
3|5 

10 

30 
60 
90 

If you operate the pull tester with the USB port not connected and if 
you are not in setting modes, the device turns off automatically after 
5 minutes without any button being pressed. Your may change this 
period by pressing the I-Button. Pressing the 0-Button you change to 
the next submenu. (Applies for FMT-W30 only) 

 

 

 

 

With a short press on the S-Button you return to the main setting menu.. 

 

 

 

 

The pull tester tares and resets the PEAK automatically with start of 
each measurement in order to eliminate any influences by change of 
the material and temperature. In some cases it might be useful to 
deactivate this functionality by pressing the I-Button  to  Select  0= 
AutoTare OFF. Pressing the 0-Button you change to the next 
submenu. 

 

 

 

 

With a short press on the S-Button you return to the main setting menu. 

 

 

 

 

If the limit comparator is activated a buzzer signalizes with an 
interrupted or permanent peep, whether the limit is not reached or 
exceeded (with FMT-W30 the USB-port has to be connected). You 
may switch OFF=0 the buzzer, although the limit comparator is 
active by pressing the I-Button. Pressing the 0-Button you return to 
the first submenu. 

 

 

 

 

With a short press on the S-Button you return to the main setting menu. 

 

 

 

 

FMT-W40 only:  
By pressing the I-Button an automatic break detection can be 
switched on and off (see chapter 3.6).  

 

 

 

 

With a short press on the S-Button you return to the main setting menu. 

 

 

 

The function code P2 can be selected in order to switch the memory 
and statistic function ON=1 and OFF=0. 

 

 

 

 


The submenu P22 configures the automatic memory function for the 
motorized pull tester. 
0 = Disable automatic memory function 
1 =  Save measuring value when reaching the limit. 
2 = Save measuring value at end of dwell time 
3 = Save peak value 

 

 

 

 

With a short press on the S-Button you return to the main setting menu. 

 

 

 

The setting menus P3 and P4 are used for limit comparator functions. (see chap. 4.1) 

 

 

 

 

 

 

 

 

 

 

To return to the originally works defaults select the setting menue 
P0. Press the 0-Button, the display shows P01 and a flashing 0. 
Press now the I-Button, until “1” is flashing. Confirm by pressing the 
S-Button that all changed setting should be resetted to the original 
defaults. The display shows now rESEt. To reconfirm that the 

μ

Processor is resetted. Leave the setting mode with a long press of 

the S-Button. 

Summary of Contents for FMT-W30C5

Page 1: ...re Terminals Dynamom tre Numerique par Jonctions de C bles Dinamom trico Digital Por cables con unions Adecuadas Dinamometri Digitales a di Cavi con Connettori BDA Version 2 4 Bedienungsanleitung Oper...

Page 2: ...l tester FMT W30 20 3 6 Start measuring AutoTara with motorized pull tester FMT W40 21 3 7 Automatic memorizing of measured values with motorized pull tester FMT W40 21 3 8 Display statistical data 22...

Page 3: ...formation 2 1 Scope of Supply unpacking and Setting Up The scope of supply consists of Measuring instrument with integrated electronics Hand lever or roller grip USB Interface cable for optional softw...

Page 4: ...ecommended date of next calibration and the nominal measuring range Fn To turn the instrument off you may press the I Button for 2 seconds else the AUTO OFF function will turn the instrument off autom...

Page 5: ...he cable connector stays securely inside the ring and cannot be pulled out through the slot The wire clamp fixture should be positioned in the full open position The manual pull tester needs a free le...

Page 6: ...e develops which works upon the cable connection The peak value which mostly appears before the connection snaps up is captured at 1000Hz and is indicated in the display After 5 seconds of unchanged m...

Page 7: ...tive if 25 of the force limit minimum tensile strength P3 or a force of 40N is reached If the actual measured force falls below this level the cable clamp returns to its home position and the measurin...

Page 8: ...etting or load new test limits The memory and static function can be turned off see chap 6 0 2S 4 Tensile force limit control Depending on the wire strength and the related standard the following tens...

Page 9: ...not apply with the motorized version FMT W40 The lower limit is adjusted automatically to 25 or max 40N of the upper limit With a short press on the S Button you return to the main setting menu I 0 Pr...

Page 10: ...icense file has to be generated and saved in the Alluris folder Please enter the received license key and the serial no of your pull tester and click generate license file The license key has to match...

Page 11: ...standard requiring 30 N the instrument is parametrized to 36 N This allows pulling further while testing than the limit set to compensate for a relaxing of the test object The test report refers to t...

Page 12: ...mpany details with your own logo etc The template is filed in C Alluris FMT W_Connect Formular xlsx Fields with gray background are filled by the data entered in FMT W_Connect and the data of the test...

Page 13: ...to the main setting menu 0 I 1 0 The pull tester tares and resets the PEAK automatically with start of each measurement in order to eliminate any influences by change of the material and temperature I...

Page 14: ...a PC 24 VDC 3A universal power supply Standby consumption 42mA Average consumption drive active 0 8A Interface USB 2 0 Optional software FMT W_Connect Temperature range Operation 5 40 C 5 30 C Storage...

Page 15: ...29 8 2 Register for updates In order to receive information about product changes and updates and to enjoy full warranty please register your product at www alluris de service php...

Page 16: ...service A loud vibrating sound occurs but the clamping device does not turn and the pull force does not increase They drive might be blocked and overloaded Switch the instrument off and remove the blo...

Page 17: ...irm additionally that the product complies with the requirements of the EMC Directive 2006 42 EEC In bereinstimmung mit der WEEE Richtlinie 2002 96 EC ist dieses Ger t eingestuft als Monitoring and Co...

Page 18: ...nischen Bundesanstalt PTB Braunschweig im Rahmen des DAkkS zur ckf hren A 2 Calibration Confirmation acc DIN EN 10204 2 1 We hereby confirm in accordance to DIN EN 10204 2 1 that this instrument has b...

Page 19: ...75 Alluris GmbH Co KG Basler Strasse 63 65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44...

Page 20: ...65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44 BDA_FMT W Art Nr 50337 Stand 07 15 Alluris GmbH Co KG Alle Rechte vorbehalten Alluris ist eine ei...

Reviews: