27
6. General settings (P1), Setting of memory (P2) and reset works defaults (Po)
To customize the instrument to the application, some of the factory settings (defaults) can be changed. These adjustments will be saved in the
microprocessor and be recalled when you switch on the force gauge the next time. In order to change the setting you press the S-Button for app.
2 second until the code P1 is displayed. Press the I-Button to choose the submenu.
S
0
I
N
kg
lb
In menu P11 you can select the measuring unit by pressing the I-
Button. The chosen unit flashes in the display. Default setting is the
SI-dimension Newton (N). Pressing the 0-Button you change to the
next submenu.
With a short press on the S-Button you return to the main setting menu.
0
I
1
2
3
5
10
The load cell captured the force with a rate of 3,6kHz. The display
update is limited to 10Hz in order to make the display readable. You
may reduce this rate by selecting an appropriate value with the I-
Button. Pressing the 0-Button you change to the next submenu.
With a short press on the S-Button you return to the main setting menu.
0
I
5
10
20
30
The measurement will be stopped automatically if no new force
values are captured for 5 seconds. You may change this period by
pressing the I-Button. Pressing the 0-Button you change to the next
submenu. (Applies for FMT-W30 only)
With a short press on the S-Button you return to the main setting menu.
0
I
1|2
3|5
10
30
60
90
If you operate the pull tester with the USB port not connected and if
you are not in setting modes, the device turns off automatically after
5 minutes without any button being pressed. Your may change this
period by pressing the I-Button. Pressing the 0-Button you change to
the next submenu. (Applies for FMT-W30 only)
With a short press on the S-Button you return to the main setting menu..
0
I
1
0
The pull tester tares and resets the PEAK automatically with start of
each measurement in order to eliminate any influences by change of
the material and temperature. In some cases it might be useful to
deactivate this functionality by pressing the I-Button to Select 0=
AutoTare OFF. Pressing the 0-Button you change to the next
submenu.
With a short press on the S-Button you return to the main setting menu.
0
I
1
0
If the limit comparator is activated a buzzer signalizes with an
interrupted or permanent peep, whether the limit is not reached or
exceeded (with FMT-W30 the USB-port has to be connected). You
may switch OFF=0 the buzzer, although the limit comparator is
active by pressing the I-Button. Pressing the 0-Button you return to
the first submenu.
With a short press on the S-Button you return to the main setting menu.
0
I
1
0
FMT-W40 only:
By pressing the I-Button an automatic break detection can be
switched on and off (see chapter 3.6).
With a short press on the S-Button you return to the main setting menu.
I
0
I
1
0
The function code P2 can be selected in order to switch the memory
and statistic function ON=1 and OFF=0.
0
I
0
1
2
3
The submenu P22 configures the automatic memory function for the
motorized pull tester.
0 = Disable automatic memory function
1 = Save measuring value when reaching the limit.
2 = Save measuring value at end of dwell time
3 = Save peak value
With a short press on the S-Button you return to the main setting menu.
I
The setting menus P3 and P4 are used for limit comparator functions. (see chap. 4.1)
I
I
0
I
0
1
To return to the originally works defaults select the setting menue
P0. Press the 0-Button, the display shows P01 and a flashing 0.
Press now the I-Button, until “1” is flashing. Confirm by pressing the
S-Button that all changed setting should be resetted to the original
defaults. The display shows now rESEt. To reconfirm that the
μ
Processor is resetted. Leave the setting mode with a long press of
the S-Button.