9.0 Frequent questions (FAQ)
The display shows numbers upside down!
The display direction changes automatically according to the posi-
tion of the device when the instrument is switched on.
Display shows values only in compressive or tensile force direction! Change the mode of operation using the S-key. The actual force
value, either tensile or compression force indicated if the Peak
symbol flashes or this symbol is not shown
3.3 Selecting the operation mode)
Force values are shown with negative prefix!
Depending on your usage you may set a negative prefix either for
tensile forces (default) or for compressive forces
ter 4.2.2 Ch/- prefix for push or pull operations (P12))
.
The depicted peak-value is higher than the maximum value!
T
he maximum value equals the average value of the last taken sing-
le values, whereas the amount of single values depends on the set
display update time. The peak value, in contrast, is the absolute
highest value measured.
Does the instrument work also after a being stored in the dark for a
longer period?
Yes. Even after a self discharging of the internal energy storage the
instrument can be used immediately. A back-up battery guarantees
the basic function of the device
.
USB communication is disturbed!
Is the right device driver installed? Devices series FMI-S must be
turned on only after connecting via USB, else they can not be de-
tected. Alternativly, press any button on the device.
The display shows “OvErL“!
Remove the excessive load immediately and check the instrument.
If the indicated measuring values are obviously untrue the load cell
has to be changed by the manufacturer.
Important note: After over-
loading the device the calibration certificate becomes invalid.
The display shows “bat.1” or “bat.0“ and the device does not res-
pond!
Press the O-button and charge the battery
Software FMI Analyze: Display shows that S-key and / or external
switches are not active (green Button) although the unit is connec-
ted!
Activate the memory function
(see p. 20 chapter 4.4.2 Setting
memory and statistics functions)
The device turns off automatically!
The auto-off function turns off the instrument, if no button has
been pressed for more than five minutes. This period of time can
be changed and set beween 1 und 90 minutes
4.2.6 Auto-Off function (instrument off) (P15))
.