6.0  Test stand application of force gauges

The rugged aluminium dye cast housing of the force gauges enables 
the use in test stands. Appropriate tapped holes for mounting are at 
the rear of the housing.

Please pay attention on that force effects are carried out perpendicu-
larly to the measuring axle.  To align the instrument properly there are 
two holes for locating pins D = 3 (t = 3) and one hole D = 8 (t = 8). To 
mount the device firmly to the stand there are two tapped holes M4 
(t = 6)  and two M5 (t = 8). Please consider the tap hole depth of the 
screws.

If the force gauge is mounted head down to the test stand, the display 
adjusts automatically with a 180° reverse on turning on the instru-
ment.

Dimensional drawings

 and 3D-CAD-data for test stand application of 

the instruments are ready for 

download

 on our website.

Manual or motor driven test stand for precise linear hub movements 
for compression or tensile force tests are a useful addition to force 
gauges. 

They allow to detect material strength in composed materials, 

operating forces of control elements or adhesive forces of packaging ma

-

terials or foils exacly and reproducible.

Please do not hesitate to contact us for more comprehensive appli-
cation advise. Product information for test stand and entire force test 
systems can also be found on our websit

www.alluris.de.

Force gauges in test stands

Force gauge FMI-B50 mounted to motor driven test stand FMT-220

3h7

M4 t=6

M5 t=6

M4 t=6

10

75

75

10

M5 t=6

3h7

8h7

57,10

46,70

40

82

150

55

13

430 g

FMI-S/B-WE_V01

FMI- S/B

WEIGHT: 

Aluminium

A1

1 OF 1

DWG NO.

TITLE:

MATERIAL:

DRAWN

NAME

SIGNATURE

DATE

ALLURIS

Rommerskirchen

26.09.2014

© bei Übermittlung schutzfähiger Informationen behält sich die 

Alluris GmbH & Co. KG, Baslerstr. 65, 

79100 Freiburg - Germany

alle Rechte für den Fall der Patentierung vor. 

Jede Verfügungsbefugnis, wie Kopier- und Weitergaberecht, bei uns.

© In case of communicating patentable information 

all rights arising in the event of a patent grant remain with

Alluris GmbH & Co. KG, Baslerstr. 65, 

79100 Freiburg - Germany.

All rights of disposition, particularly for 

copying or distributing, are reserved.

The English translation is believed to be accurate. 

In case of discrepancies the german version shall govern.

1

2

3

4

1

2

3

A

B

C

D

E

A

B

C

D

E

innovative Messinsrtumente

             für physikalische Größen

_

_

Dimensional drawing  
(download click on image)

Summary of Contents for FMI-B Series

Page 1: ...ucciones de servcio Istruzioni per l uso Alluris GmbH Co KG Subject to changes and corrections without notice 2016 04 13_EN Operation Manual Force Gauges FMI S FMI B FMI B FMI S Deutsche Version Engli...

Page 2: ...S P14 17 4 2 5 Automatic measuring stop Auto Stop for FMI B P14 18 4 2 6 Auto Off function instrument off P15 18 4 2 7 Automatic taring function Auto Tara P16 18 4 3 Resetting factory defaults PO 19 4...

Page 3: ...ay be can destroy the internal sensor Please consider the technical data in regard to the environmental conditions see p 30 chapter 7 0 Technical Data The instrument is equipped with a temperature com...

Page 4: ...s de 2 1 Scope of supply Standard scope of delivery part no o Optional accessories not included FMI B10 FMI S10 FMI B20 FMI S20 FMI B30 FMI S30 FMI B50 FMI S50 Protection case Base unit Type label wit...

Page 5: ...the instrument for more than 2 month in the dark you should expose it to light for at least 8 hours in the regu lar working environment 250 Lux before use Or you may use the USB cable to charge the i...

Page 6: ...SO Motor control Drive unitcontrol FMT 220 MK USB 2 0 Mini B socket Function Accessory part Part number Data transfer to MS Excel1 Software FMI_Connect FMI 972 Data analysis on PC Software FMI_Analyze...

Page 7: ...after long storage in the dark 6 10 pole Hirose socket series FMI B30 und FMI S30 and higher for digital I O s service calibration etc 7 I key On Off press 2 seconds Start Stop measurement select next...

Page 8: ...not use tools to screw adapters onto the instrument s threaded axle Radial and side forces can da mage the instrument As gauges are often used in the scope of destruction tests or for appraising secur...

Page 9: ...on instrument off P15 Force gauges with LiPo battery series FMI B measure upon each start the remaining capacity If it is less than about 4 hours bat 2 is dis played on the screen In this case charge...

Page 10: ...pressing the S key 3 3 1 Standard mode In standard mode the actual measured value is displayed By default the display refresh rate is 10 Hz In the settings menue P13 you may change the refresh rate s...

Page 11: ...ent indicates the peak value for tension operations You may change the prefix see p 17 chapter 4 2 2 Changing prefix for push or pull operations P12 Instruments series FMI B20 and FMI S20 and higher c...

Page 12: ...top function see p 17 chapter 4 2 4 Automatic measu ring stop Auto Stop for FMI S P14 allows to stop a measurement automatically if measured values remain stable within a defined period of time The fe...

Page 13: ...shows the mea sured real time value In all of the three peak modes the instrument captures and stores beside the actual measured value also the posi tive and negative peak value In Peak mode or Peak...

Page 14: ...d values can be deleted one after the other by pressing the O key The display shows O O after having deleted a value By pressing the S key you can select the next stored value and proceed in the same...

Page 15: ...c settings menus P1 General settings see p 16 chapter 4 2 General settings P1 P2 Memory FMI B30 and FMI S30 and higher see p 19 chapter 4 4 Memory and statistic functions P2 P3 Upper limit FMI B30 and...

Page 16: ...enu by selecting the next submenu pressing the O key or by reverting to the menu on the level above By pressing the S key you revert to the level above By pressing the S key for 2 seconds you revert t...

Page 17: ...ency of 3 6 kHz To ensure the legibility of the display the display refresh rate is limited by default to 10 Hz Submenu P13 allows to reduce this value further Press the I key to select the desired fr...

Page 18: ...evel 4 2 6 Auto Off function instrument off P15 Force gauges running without USB cable are turned off if no button is pressed for more than five minutes as long as you are not working in the settings...

Page 19: ...S key Reset factory defaults 0 no Reset 1 Reset 4 4 Memory and statistic functions P2 4 4 1 General considerations on memory functions Force gauges series FMI B30 and FMI S30 and higher are able to c...

Page 20: ...es Series Pressing the S key for 2 seconds the instrument returns to the opera tion level in idle status To confirm activated memory the display now shows the according Mem symbol in addition to the s...

Page 21: ...individual values You may pause the recording of measurement values while the measuring keeps running and restart the recording both by pressing the S key Recording ends with the termination of the me...

Page 22: ...djacent chart Standard mode is set last measured value S S S S S S S S S S N N Mem Peak modes are set Peak last measured value Mem Peak N N N Peak positive peak value N N Mem Peak negative peak value...

Page 23: ...hen by software Devices of series FMI S have to be connected to PC already during measurements 4 4 7 Clearing memory You have three possibilities to clear the memory Single peak values can be deleted...

Page 24: ...As soon as the measured value is smaller than the lower limit value a downward pointing arrow is shown on the right side of the display Accordingly an upward pointing arrow is shown if the measured v...

Page 25: ...oad or the exceeding of limit values via USB 2 0 or Hirose cable to a PC or a motor driven test stand 5 1 Data transfer via USB For data tranfer via USB you need a USB cable part no FMI 931USB and a s...

Page 26: ...FMI 976 offers basically all analyzing functions as FMI_Analyze Furthermore the software packa ge allows to control the optional motor for test stands series FMT 220 and enables the setting of parame...

Page 27: ...or 100 mm FDM 250C1 with a distance of 25 mm respectively 45 mm to the reference mark 5 2 Hirose Socket Force gauges series FMI B30 and FMI S30 and higher have a 10 pole Hirose socket for several ext...

Page 28: ...GND 3 Gr n Green DIGIN1 Digital input 4 Gelb Yellow TRIGGER_OUT C Trigger signal measuring frequency 5 Grau Grey TRIGGER_OUT E 6 Orange Orange DIGOUT1 C Digital output setting via FMI_Analyze 7 Blau B...

Page 29: ...sed materials operating forces of control elements or adhesive forces of packaging ma terials or foils exacly and reproducible Please do not hesitate to contact us for more comprehensive appli cation...

Page 30: ...0 15 Temperature offset Tk relative K 0 02 O Point offset To absolute K 0 02 Permissible error of measurement 0 5 0 3 Tracking To surpressed Auto Tara automatic compensation Overload output USB interf...

Page 31: ...xcluded from this warranty Besides the warranties as agreed upon in our general terms and conditions shall apply 8 2 Product registration In order to receive information about product changes and upda...

Page 32: ...he dark for a longer period Yes Even after a self discharging of the internal energy storage the instrument can be used immediately A back up battery guarantees the basic function of the device see p...

Page 33: ...protection laws would have to be considered as free A 2 Calibration confirmation acc DIN EN 10204 2 1 We hereby certify that the equipment has been tested in the production process in accordance with...

Page 34: ...79100 Freiburg Germany Type Number FMI S10xx FMI S20xx FMI S30xx FMI S50xx FMI B10xx FMI B20xx FMI B30xx FMI B50xx see type label Description Digital Force Gauge Serial number See type label Applicab...

Reviews: