3.4. Start/Stop Measurements
Switch the instrument on and bring it into the measurement position.
Start:
Start the measurement by pressing the I-key. The display shows „O.O“.
Measure:
The display shows the current measured value or the peak value
p. 10 chapter 3.3 Selecting the operation mode)
. The symbol of the
measuring dimension flashes as long as the instrument is measuring.
Stop:
By pressing the I-key the measuring cycle is completed.
After stopping the measurement, the display shows the measured
value. Depending on the selected operation mode, this is either the
last measured value or the respective peak value. The dimension unit
is now displayed continuously without flashing.
The auto-stop-function
(see p. 17 chapter 4.2.4 Automatic measu-
ring stop – Auto-Stop for FMI-S (P14))
allows to stop a measurement
automatically, if measured values remain stable within a defined
period of time. The feature can be deactivated if the instrument is
connected to an external power source.
Please make sure that force – both compressive and tensile forces – is
always applied perpendicularly to the measuring axle. The measuring
results will be incorrect if the force axle is not aligned perpendicularly.
Forces that are not applied perpendicularly can damage the internal
sensor especially of gauges with a measurement range designed for
minute forces, when the radial or laterally operating forces are too big.
N
N
idle
start
running measurement:
stop
result
N
N
N
N
unit symbol flashes
Indications on display during measurements
90°
500 N
max
.
St
op
<>
Pe
ak
Dat
a
Unit
Cle
ar
S
St
ar
t
>
<
O
S
<>
disp
lac
emen
t gauge
FMI-
S50
F
or
ce ++
Keep the instrument always perpendicular to the probe
Correct positioning of the instrument