Auto-Pure series Operation Manual
Chapter 2 Specification
─
2
─
Chapter 2 Specification
1. Normal operating conditions
Environmental Temperature: 10
C
~
35
C
Relative Humidity ≤70%
Input: AC 100
~
240V
,
50Hz/60Hz
2. Basic parameter & Performance
Table 1 Basic parameter & Performance
Model
Parameter
Auto-Pure
32A
Auto-Pure
48A
Auto-Pure
20A
Auto-Pure
20B
Auto-Pure
24D
Auto-Pure
24BT
Principle
Magnetic Particle Method
Throughput
1~32
1~48
1~20
1~20
1~24
1~24
Kits
96
Deep-well
plate
96
Deep-well
plate
2ml Kits
5ml Kits
10ml Kits+
2ml tubes
5ml Kits+
2ml tubes
Sample
Volume/μL
50~1000
50~1000
50~3000
50~5000
50~10000
50~5000
Stability
CV≤5%
Lysis temp.
Ambient temperature ~120
℃
Elution temp.
Ambient temperature ~120
℃
Heating time
Heating time
(
Ambient temperature ~120
℃)≤
4 minutes
Temperature
Accuracy
±1
℃
Operation
interface
7-inch touch screen, 3 shortcut buttons and mouse is available
Built-in protocol
8 groups of preset protocols, 100 groups of protocols can be stored
Protocol
management
New, Edit, Delete, Save as
Expansion
interface
Standard USB, ethernet port and WIFI are available
Network
Extensible Ethernet remote control, WiFi, 4G network
Pollution control
UV light
Exhaust way
By Fan
Data storage
Available, with built-in SD card
Max.input
power
450W