25
4. Common Alarm Code Indication
AMC2100 Alarm Table
Code
Contents Meaning
Solution
8
OV. Flow
Over range
Increase range
16
OV. Flow
Exceeding AD value of flowrate
Slow fluid velocity
24
OV. Flow
Over range
Exceeding AD value of flowrate
Increase range
Slow fluid velocity
32
OV. Freq
Exceeding Fmax range
Increase the frequency setting range
64
UPPER
Exceeding the upper limit value of
alarm settings
Turn off alarm output or raise alarm upper limit
value
72
OV. Flow ; UPPER
Exceeding the upper limit value of
alarm settings
Over range
Increase range
Turn off alarm output or raise alarm upper limit
value
128
LOWER
Under the lower limit value of alarm
setting
Turn off alarm output or decrease the alarm
lower limit value
136
OV. Flow ; LOWER
Over range
Under the lower limit value of alarm
setting
Increase range
Turn off alarm output or decrease the alarm
lower limit value
256
EMPTY
Empty pipe alarm
Check the pipe & full the pipe
512
Zero
Zero mv value>99mv
Re-zero when fluid is static.
1024
Coil
Exciting Current<40%
Check the XY wiring fall off or not
1032
OV. Flow ; Coil
Over range
Exciting Current <40%
Increase range
Check the XY wiring fall off or not
1088
UPPER ; Coil
Exceeding the upper limit value of
alarm settings
Exciting Current <40%
Turn off alarm output or raise alarm upper limit
value
Check the XY wiring fall off or not
1152
LOWER ; Coil
Under the lower limit value of alarm
setting
Exciting Current <40%
Turn off alarm output or decrease the alarm
lower limit value
Check the XY wiring fall off or not
1280
EMPTY ; Coil
Empty pipe alarm
Exciting Current <40%
Check the pipe to full the pipe
Check the XY wiring fall off or not
1536
Zero; Coil
Zero mv value>99mv
Exciting Current <40%
Re-zero when fluid is static.
Check the XY wiring fall off or not