Alia AMC2100 Series Operation Manual Download Page 11

 
 

 10

3.5 Auto zero Adjustment 

When the tube are full pipe, and no flow, flowmeter still show the flowrate, then you could use Auto Zero adjustment to 
adjust your flowmeter to Zero, the method as follow: 

  
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

3.6 Check for power off message 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

In this setting, Press 

 or 

 button till you selected ZERO and press 

ENT

 to confirm 

 
 

    OPERATIONS 

 

SET    

 ZERO 

   ESC 

 

In the normal display, press 

ENT

 button for 5 seconds, then it will go to next setting 

 
 

 



123.456 m3 

+0.1234

m3/h 

    

    

  

  

 

   

  

 

Select  YES, and press 

ENT

 to confirm . 

 
 

           ZERO 

 

NO          

YES

 

 

ENT

 

ENT

 

ENT

 



123.456 m3 

+0.0000

m3/h 

    

    

 

 

  

  

 

In general display, Press 

 or 

 button to select   

Press ENT to confirm, then check for the 

power off message window 
 
 

Power off message window 

ON: date & time when power on                 OFF: date & time when power off                 
Q: the flowrate  when power off



the totalizer  when power off 

Example: 01 means the power off recored series No.(It could record 99 times of power off) 
Press PgUP or PgDN   check the history record of power off. 

Ps: Press PgDN till the last page, OFF option means current date, time, and flowrate.  

 

ON             

2008/04/23

 

         

                 13:30:55 

OFF          

 

2008/04/23

 

                       

13:30:55   

Q:          1.0234 m3/h 



4.123  m3

01

  PgUP   PgDN  ESC

 

ENT

 

Zero Adjustment.  
 
 

 
 Zero    Trimming … 
 
 

Zero Adjustment  end. Select ESC,  press 

ENT

 to turn back to the operations select 

(If you want to do Zero Adjustment again, Pls select ZERO ) 
 

Zero   Tirm  End 
 -2.624mv 
 
Zero                    Esc 

ENT

 

Select ESC,  press 

ENT

 to confirm and  turn back to the normal display 

 
 

    OPERATIONS 

 

SET     ZERO   

 ESC

 

 

ENT

 

the normal display 

 
 



123.456 m3 

+0.0000

m3/h 

    

    

  

   

     

ENT

 

Summary of Contents for AMC2100 Series

Page 1: ...ALIA TECHNOLOGY LLC...

Page 2: ...verter AMC2100 Operate Manual 7 3 1 Converter AMC2100 Operate Manual 7 3 1 1 Power and signal output connected line compact and separate model 7 3 1 2 Separate Wiring 8 3 1 3 AMC2100 Panel deploy 8 3...

Page 3: ...3 5 15 95 155 3 5 20 105 160 4 5 25 115 166 4 5 32 140 180 6 5 40 150 190 7 0 50 165 201 9 5 65 185 220 12 80 200 235 15 100 220 254 17 125 250 284 21 150 300 285 314 28 200 350 340 369 36 250 400 395...

Page 4: ...tion Wafer 1 3 AMF301 Installation 1 4 AMF500 Installation mm Inch L D 25 1 90 71 32 1 1 4 80 40 1 1 2 86 50 2 115 100 65 2 1 2 115 120 80 3 130 131 100 4 155 151 125 5 155 181 150 6 185 206 200 8 215...

Page 5: ...nge should be flat having no sharp residue Otherwise liner will be damaged After flowmeter is installed if other places in pipe needs to be welded flowmeter s power must be shut down 2 Usually there w...

Page 6: ...and downstream of the installation should satisfied the following conditions Picture Generally tube installation the flowmeter Upstream could try to stay 5D 10D Downstream 2D 5D 2 4 Install method a...

Page 7: ...he flowmeter tube it may cause Liner inside broken Please DO NOT use your hand or rope to hang on Converter or wiring box if Flowmeter size is bigger than 80mm because the material of converter and wi...

Page 8: ...cide if you need to connect according to your needs 1 20mA output 4 2 4 20m output A 3 Pulse output 4 ulse P output 5 RS485 output 6 RS485 output Cable from sensor Power supply 90 260V AC 50 60Hz or 2...

Page 9: ...level Save presently settled parameter Jump to next parameter setting Up Selected one of 4 display line Change present digit dot and survey unit Up Down movement Right Change the selected line on the...

Page 10: ...u need Line 1 display from m3 h turn to 3 4 Totalizer Reset 123 456 m3 0 0000m3 h 123 456 m3 0 0000m3 h In normal display press button until Line 2 is selected 123 456 m3 0 0000 FS Press ENT button to...

Page 11: ...f message window Power off message window ON date time when power on OFF date time when power off Q the flowrate when power off the totalizer when power off Example 01 means the power off recored seri...

Page 12: ...mping s Hydro State Baudrate Data Bit Parity Stop Bit Modbus Address 0000 Direction MODBUS Type 123 456 m3 0 0000m3 h OPERATIONS SET ZERO ESC Empty Detect Empty Limit Communication ENT ENT ENT ENT ENT...

Page 13: ...nsor Factor ENT ENT ENT ENT ENT Convertor Factor Coil Current A Restore Shipping Convertor NO ENT ENT ENT Loop Test Pulse Test Coil Test Password Coil Freq Hz Zero Factor ENT ENT ENT ENT 4mA Trim ENT...

Page 14: ...ress or button to selected the output 1 Default 2 Pulse 3 Frequency 4 Alarm Pulse output Each pulse match the Liter value Liter Pulse Press button to move Press to changed the Value Pulse width Press...

Page 15: ...if it needs empty pipe detect Empty pipe detect the value Press button to move Press to change the value Max Limit fs 05 Response speed adjustment Press button to move Press button to change the value...

Page 16: ...Press button to move Press button to change the digit TIME hh mm ss 17 09 49 Time Setting Press button to move Press button to change the digit Pls press ENT button for 5 seconds then it will go back...

Page 17: ...ART Address 00 DATE yy mm dd 08 01 10 Date Setting Press button to move Press button to change the digit ENT TIME hh mm ss 17 09 49 Time Setting Press button to move Press button to change the digit P...

Page 18: ...tton to select the simulated output value 4 8 12 16 20 Password 1111 Password login window Input the system password Password 1111 OPERATIONS SET ZERO ESC ENT Pulse Simulated output Press or button to...

Page 19: ...ose NO to skip to Restore Factory Settings choose YES to conduct Full Tube Trim If there is full tube empirical value you can choose Keyin to modify full tube frequency value Otherwise it is not recom...

Page 20: ...ess or button to select the calibration type 4 8 12 16 20 Pulse Simulated output Press or button to select the calibration type 1 10 100 1000 2000 5000 Excit voltage calibration Check the coil X to Y...

Page 21: ...ess button to move Press to change the value Pls press ENT button for 5 seconds then it will go back to the normal display Restore the parameter to second memory Press or button to select if it needs...

Page 22: ...Revise valve 0 943 Point 2 1 0 983 1 017 new Revise valve 1 017 Point 3 2 2 046 0 978 new Revise valve 0 978 Point 4 4 4 176 0 958 new Revise valve 0 958 Filled the 4 points Revise valve in order to...

Page 23: ...m3 h the actual speed was 5 66 m s display 4012 m3 h Velocity 5 677 m s New Revise 4000 4012 0 997 Flow Velocity 5 66 m s during Revise_5 4 m s and Revise_6 6 m s so you can set as follow Speed_5 5 66...

Page 24: ...lected SET Alarm Select MAX MIN BAT DIR Alarm Select Press or button to BAT 1 Max Upper Alarm 2 Min Lower Alarm 3 Batch 4 Direction ENT Low Cutoff fs 0 0 Function for cutoff the low flowrate Press but...

Page 25: ...on to AUTO or MANUAL ENT SET 1 000m3 RUN 0 000m3 START CLR EXIT Press or button to START To start batch control If choose CLR The actual flow value will be zero ENT SET 1 000m3 RUN 0 000m3 PAUSE In th...

Page 26: ...lue of alarm setting Increase range Turn off alarm output or decrease the alarm lower limit value 256 EMPTY Empty pipe alarm Check the pipe full the pipe 512 Zero Zero mv value 99mv Re zero when fluid...

Page 27: ...26 633 W 5th Street 26th Floor Los Angeles CA 90071 USA Tel 1 213 533 4139 Fax 1 213 223 2317 URL www alia inc com Email alia inc com alia...

Reviews: