background image

 
Page  11 of  23 

USER MANUAL 

ADS intelligent 

 

 

 

 

Softwareversion: 

0.95 

 
 
 
 

 

9.2

 

Alarms-and warnings 

 

 

 

 

 
Inside  this  menu  the  patient  movement  detection  (PMD)  and  the  output  of  alarms  can  be  set 
up.  By  number  one  (1)  the  operator  can  choose  between  "no  patient  movement  detection",  
"patient  movement  detection  within  30  minutes  ",  patient  movement  detection  within  60 
minutes"  and "patient  movement  detection  within  90 minutes".   
By  leaving  the  submenu  by  pressing  the  "√"  button  in  the  lower  right  corner  all  changes  will 
be saved.    

9.2.1

 

PMD and  alternating/change  pressure cycle mode 

When  the  operator  selects  the  PMD  time  of    30,  60  or  90  minutes,  the  ADS  intelligent  starts 
the  patient  movement  monitor.  The  selected  time  the  device  will  record  and  notice  or 
movements  of  the  patient.  The  time  passed  by  with  "no  movement"  will  be  shown  in  the  bar 
below  the  patient  movement  monitor.  The  white  part  of  the  bar  shows  the  time  in  percent  of 
the  choosen  monitoring  time  no detected movement  above the necessary  movement  level.   
After  every  detected  movement  about  the  necessary  level,  the  counter  will  be  reset  and  the 
movement  detection  starts  from  zero  minutes  again.   
Wilmslow  movement  be  recorded  and  detected  within  the  chosen  time  and  the  chosen  total 
time  has  completely  passed by, the device  will  react in  the  selected  way of  alarm  setting: 
 
Alternating/change  pressure  cycle  mode NOT activated:    the  system  will  give  alarm   
 
Alternating/change  pressure  cycle  mode activated:   
 
The  system  will  give  warning  "  no  movement"  and  will  turn  into  the  "alternating/change 
pressure  cycle  mode"  for  exactly  one  time.  After  that  the  second  period  of  movement 
recording  with  the  chosen  period  of  time  will  start  again.  If  a  significant  moving  above  the 
necessary  level  will  be  recorded  within  this  second  period  of  time,  the  warning  in  the  display 
will  be delete  the  system  starts  again  with  the  normal  patient  movement  monitoring. 
If  NO  significant  moving  above  the  necessary  level  will  be  recorded  within  this  second  period 
of time,  the system  will  give  "alarm"  at the  end of the  chosen  monitoring  time.     

Summary of Contents for ADS Intelligent

Page 1: ...5 7 Turning on the device 7 8 Screen the beginning after turning on the ADS intelligent 8 9 Settings 10 9 1 Weight setting 10 9 2 Alarms and warnings 11 9 2 1 PMD and alternating change pressure cycle...

Page 2: ...he corresponding mattress can be used till pressure ulcers gradation 4 2 Operation the device Connecting the device to the power supply The ADS intelligent must connect to a power strip of 230 V 50Hz...

Page 3: ...ect the mattress and the bed using the top straps Otherwise there is danger that the maneuvarability of the bed and even the functionality of bed will be limited In addition there is a danger of damag...

Page 4: ...em is operational The patient may not guaranteed bed sores by pressure relief at the filling process and missing volume fill early The ADS intelligent features a log system that duration the error cod...

Page 5: ...tests for electrical safety for electrical devices Annual maintenance is recommended because the ADS intelligent is a high quality medical device The product life cycle can be extended in case of regu...

Page 6: ...Page 6 of 23 USER MANUAL ADS intelligent Softwareversion 0 95...

Page 7: ...USER MANUAL ADS intelligent Softwareversion 0 95 7 Turning on the device After driving the main switch in position I the AirMed PLUS Logo and software version of the device is shown for approximately...

Page 8: ...the given value pressure inside the cell is lower than the given value 2 This symbol shows that choosing patient s weight class 3 This symbol shows to enter the settings 4 Alarm signal It will be show...

Page 9: ...ing please check chapter 9 2 Remark one the patient movement monitor can only be used in static therapeutic mode because the mode of alternating change pressure cycle the movement detection is not pos...

Page 10: ...enu 9 1 Weight setting In this menu you can adjust the pressure inside the cells according to the patient s weight Press for increasing and press for decreasing There are 10 possible weight steps avai...

Page 11: ...time no detected movement above the necessary movement level After every detected movement about the necessary level the counter will be reset and the movement detection starts from zero minutes agai...

Page 12: ...Warnings In opposite to the alarm warnings will be only shown visually and not acoustically The number three 3 warnings can be configured also as acoustically or nurse call system Acoustic if enabled...

Page 13: ...adjusted 9 3 2 Logic nurse call by this button the nurse call logic hardware relais can be permanently turned one or off If the nurse call logic is deactivated the relais will be activated in the case...

Page 14: ...d Time 9 5 Date setup Please see chapter 10 Set up Date and Time 9 6 Touch calibration by using this function the touch screen will be calibrated The operator needs to the crosses in the display withi...

Page 15: ...Page 15 of 23 USER MANUAL ADS intelligent Softwareversion 0 95 Inside this submenu the operator can select one language...

Page 16: ...mode 9 8 1 Bedding mode In this mode the mattress will be inflated with full pressure for maximum 20 minutes After 20 minutes the system turns into the previous mode automatically again 9 8 2 head en...

Page 17: ...s shown in the display and in the schematic mattress is marked by a small red bar above the motion monitor graph 1 By touching the time graph the operator can scroll horizontally from early to late in...

Page 18: ...e current figure which is changed is highlighted in black colour 2 By pressing this button 2 the new settings will be saved 3 By pressing this button the operator will leave this submenu without savin...

Page 19: ...patient monitoring a without redistributing any movement of patient please see chapter Error Reference source not found for details AND b Alternating cycle time mode is enabled One cycle of alternati...

Page 20: ...le format and using the data The data will be saved in CSV format and can be opened at the PC by Excel of Calc One file per day will be saved with the name year month day csv Example a data file with...

Page 21: ...Alarme sind m glich Warnings Priority Pump up 1 The mattress is still into pump up procedure and it is not allowed to place the patient onto the mattress If the pressure inside each cell is sufficient...

Page 22: ...following medical device Product name ADS intelligent Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with risk class 1 bas...

Page 23: ...USER MANUAL ADS intelligent Softwareversion 0 95 manufacturer of ADS intelligent AirMed PLUS GmbH Harpener Feld 34 D 44805 Bochum Germany Tel 0049 0 234 91773320 Fax 0049 0 234 91773322 Email service...

Reviews: