Alarms  and Warnings 

 
Beside  the  optical  symbols  of  alarms,  alarms  will  be  indicated  also  acoustically  (and  if 
enabled  by  the  nurse  call  system).  If  an  acoustic  alarm  will  come  up,  you  can  see  a  bell  into 
the  display  as well.   
The  alarm  sound  will  be  active  until  the  operator  quits  the  alarm  sounds  by  pressing  the  alarm 
key button  , which  will  be confirmed  by a signal  beep of one second.   
If  an  alarm  has  been  quitted,  the  bell  is  crossed  out  in  the  display,  but  it  is  still  shown  as  the 
alarm  signal.    
 
 
For  the  case  of  warnings:  also  warnings  can  be  put  out  acoustically  and/or  by  the  nurse  call 
system  when  it  is  enabled.  In  this  case, the warnings  can be quitted  as well.   
 
The  visual  sign  of  warning  will  also  stay  in  the  display  as  a  visible  sign  until  the  reason  for 
this  warning  or reason  for this  alarm  will  be solved.   
Alarms  have  higher  priority  than  warnings.  An alarm  over and warning  in  the  display.    
 
Folgende  Warnungen  und  Alarme  sind  möglich: 

Warnings: 

Priority 

 

Pump  up: 

The  mattress  is  still  into  pump-up-procedure  and  it  is  not 
allowed  to  place  the  patient  onto  the  mattress.  If  the 
pressure  inside  each  cell  is  sufficient  and  is  stable  for 
minimum  of  10 seconds, the  warning  will  go away.   

No Movement: 

This  warning  will  come  up  if  alternating/cycle  mode  is 
enabled  AND  within  the  chosen  patient  monitoring  time 
NO patient  movement  is  being  detected. 

Alarms: 

 

 

Leackage: 

This  alarm  will  come  up  if  within  10  minutes  the  given 
pressure  inside  the  cells  is  not  correct  and  can't  be  reached 
in  a  sufficient  value  -  in  the  meaning  of  "  pressure  is  too 
low". 

No Movement: 

This  warning  will  come  up  if  alternating/cycle  mode  is 
ENabled  AND  within  the  chosen  patient  monitoring  time 
NO  patient  movement  is  being  detected  AFTER  the  second 
run  (passed  the  chosen  patient  monitoring  time  with  "no 
movement"  for  the SECOND time).   
If  alternating/cycle  mode  is  DISabled,  this  warning  will 
come  up  when  within  the  chosen  patient  monitoring  time 
NO  patient  movement  is  being  detected  AFTER  the  FIRST 
run  (passed  the  chosen  patient  monitoring  time  with  "no 
movement"  for  the FIRST  time)   

Internal  Error: 

Fault  on PC board or µC-System 

 

Summary of Contents for ADS Intelligent

Page 1: ...5 7 Turning on the device 7 8 Screen the beginning after turning on the ADS intelligent 8 9 Settings 10 9 1 Weight setting 10 9 2 Alarms and warnings 11 9 2 1 PMD and alternating change pressure cycle...

Page 2: ...he corresponding mattress can be used till pressure ulcers gradation 4 2 Operation the device Connecting the device to the power supply The ADS intelligent must connect to a power strip of 230 V 50Hz...

Page 3: ...ect the mattress and the bed using the top straps Otherwise there is danger that the maneuvarability of the bed and even the functionality of bed will be limited In addition there is a danger of damag...

Page 4: ...em is operational The patient may not guaranteed bed sores by pressure relief at the filling process and missing volume fill early The ADS intelligent features a log system that duration the error cod...

Page 5: ...tests for electrical safety for electrical devices Annual maintenance is recommended because the ADS intelligent is a high quality medical device The product life cycle can be extended in case of regu...

Page 6: ...Page 6 of 23 USER MANUAL ADS intelligent Softwareversion 0 95...

Page 7: ...USER MANUAL ADS intelligent Softwareversion 0 95 7 Turning on the device After driving the main switch in position I the AirMed PLUS Logo and software version of the device is shown for approximately...

Page 8: ...the given value pressure inside the cell is lower than the given value 2 This symbol shows that choosing patient s weight class 3 This symbol shows to enter the settings 4 Alarm signal It will be show...

Page 9: ...ing please check chapter 9 2 Remark one the patient movement monitor can only be used in static therapeutic mode because the mode of alternating change pressure cycle the movement detection is not pos...

Page 10: ...enu 9 1 Weight setting In this menu you can adjust the pressure inside the cells according to the patient s weight Press for increasing and press for decreasing There are 10 possible weight steps avai...

Page 11: ...time no detected movement above the necessary movement level After every detected movement about the necessary level the counter will be reset and the movement detection starts from zero minutes agai...

Page 12: ...Warnings In opposite to the alarm warnings will be only shown visually and not acoustically The number three 3 warnings can be configured also as acoustically or nurse call system Acoustic if enabled...

Page 13: ...adjusted 9 3 2 Logic nurse call by this button the nurse call logic hardware relais can be permanently turned one or off If the nurse call logic is deactivated the relais will be activated in the case...

Page 14: ...d Time 9 5 Date setup Please see chapter 10 Set up Date and Time 9 6 Touch calibration by using this function the touch screen will be calibrated The operator needs to the crosses in the display withi...

Page 15: ...Page 15 of 23 USER MANUAL ADS intelligent Softwareversion 0 95 Inside this submenu the operator can select one language...

Page 16: ...mode 9 8 1 Bedding mode In this mode the mattress will be inflated with full pressure for maximum 20 minutes After 20 minutes the system turns into the previous mode automatically again 9 8 2 head en...

Page 17: ...s shown in the display and in the schematic mattress is marked by a small red bar above the motion monitor graph 1 By touching the time graph the operator can scroll horizontally from early to late in...

Page 18: ...e current figure which is changed is highlighted in black colour 2 By pressing this button 2 the new settings will be saved 3 By pressing this button the operator will leave this submenu without savin...

Page 19: ...patient monitoring a without redistributing any movement of patient please see chapter Error Reference source not found for details AND b Alternating cycle time mode is enabled One cycle of alternati...

Page 20: ...le format and using the data The data will be saved in CSV format and can be opened at the PC by Excel of Calc One file per day will be saved with the name year month day csv Example a data file with...

Page 21: ...Alarme sind m glich Warnings Priority Pump up 1 The mattress is still into pump up procedure and it is not allowed to place the patient onto the mattress If the pressure inside each cell is sufficient...

Page 22: ...following medical device Product name ADS intelligent Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with risk class 1 bas...

Page 23: ...USER MANUAL ADS intelligent Softwareversion 0 95 manufacturer of ADS intelligent AirMed PLUS GmbH Harpener Feld 34 D 44805 Bochum Germany Tel 0049 0 234 91773320 Fax 0049 0 234 91773322 Email service...

Reviews: