Page 9 of 23
USER MANUAL
ADS intelligent
Softwareversion:
0.95
peaks that higher was the causing movement. The green line shows the level which
needs to be reached to detect the movement of the patient. Will a movement be
detected, the movement monitor shows an additional red dotted line to make an
exception reporting more clear. The black bar below the movement monitor shows the
time which has passed by (white) and the remaining time (black) in % if the "patient
movement recording" is enabled. For patient movement recording please check
chapter 9.2.
Remark one: the patient movement monitor can only be used in "static therapeutic
mode", because the mode of alternating/change pressure cycle, the movement
detection is not possible because some cells are deflated. When the setting will be
done , pink-marked-line will be shown above the threshold level to make clear that the
setting has been done during this period of time.
6.
Date and time
7.
Some of bed or head-ends-up position
the current status of enter position and setting will be shown and the pressure inside
cells will be adjusted according to choosing mode:
lying: weight-referring-pressure level
head-end-up: pressure level increased by +10%
bed or nursery mode: maximal pressure level