background image

 
Page  9 of  23 

USER MANUAL 

ADS intelligent 

 

 

 

 

Softwareversion: 

0.95 

 
 
 
 

peaks  that  higher  was  the  causing  movement.  The  green  line  shows  the  level  which 
needs  to  be  reached  to  detect  the  movement  of  the  patient.  Will  a  movement  be 
detected,  the  movement  monitor  shows  an  additional  red  dotted  line  to  make  an 
exception  reporting  more  clear.  The  black  bar  below  the  movement  monitor  shows  the 
time  which  has  passed  by  (white)  and  the  remaining  time  (black)  in  %  if  the  "patient 
movement  recording"  is  enabled.  For  patient  movement  recording  please  check 
chapter  9.2. 
 

Remark  one:  the  patient  movement  monitor  can  only  be  used  in  "static  therapeutic 
mode",  because  the  mode  of  alternating/change  pressure  cycle,  the  movement 
detection  is  not  possible  because  some  cells  are  deflated.  When  the  setting  will  be 
done , pink-marked-line will be shown above the threshold level to make clear that the 
setting has been done during this period of time. 

 

 

6.

 

Date and  time 

 

7.

 

Some of   bed or head-ends-up  position   

the  current  status  of  enter  position  and  setting  will  be  shown  and  the  pressure  inside 
cells  will  be adjusted  according  to choosing  mode: 

 

lying:  weight-referring-pressure  level 

 

head-end-up:  pressure  level  increased  by +10% 

 

bed or nursery  mode:  maximal  pressure  level 

 
 

Summary of Contents for ADS Intelligent

Page 1: ...5 7 Turning on the device 7 8 Screen the beginning after turning on the ADS intelligent 8 9 Settings 10 9 1 Weight setting 10 9 2 Alarms and warnings 11 9 2 1 PMD and alternating change pressure cycle...

Page 2: ...he corresponding mattress can be used till pressure ulcers gradation 4 2 Operation the device Connecting the device to the power supply The ADS intelligent must connect to a power strip of 230 V 50Hz...

Page 3: ...ect the mattress and the bed using the top straps Otherwise there is danger that the maneuvarability of the bed and even the functionality of bed will be limited In addition there is a danger of damag...

Page 4: ...em is operational The patient may not guaranteed bed sores by pressure relief at the filling process and missing volume fill early The ADS intelligent features a log system that duration the error cod...

Page 5: ...tests for electrical safety for electrical devices Annual maintenance is recommended because the ADS intelligent is a high quality medical device The product life cycle can be extended in case of regu...

Page 6: ...Page 6 of 23 USER MANUAL ADS intelligent Softwareversion 0 95...

Page 7: ...USER MANUAL ADS intelligent Softwareversion 0 95 7 Turning on the device After driving the main switch in position I the AirMed PLUS Logo and software version of the device is shown for approximately...

Page 8: ...the given value pressure inside the cell is lower than the given value 2 This symbol shows that choosing patient s weight class 3 This symbol shows to enter the settings 4 Alarm signal It will be show...

Page 9: ...ing please check chapter 9 2 Remark one the patient movement monitor can only be used in static therapeutic mode because the mode of alternating change pressure cycle the movement detection is not pos...

Page 10: ...enu 9 1 Weight setting In this menu you can adjust the pressure inside the cells according to the patient s weight Press for increasing and press for decreasing There are 10 possible weight steps avai...

Page 11: ...time no detected movement above the necessary movement level After every detected movement about the necessary level the counter will be reset and the movement detection starts from zero minutes agai...

Page 12: ...Warnings In opposite to the alarm warnings will be only shown visually and not acoustically The number three 3 warnings can be configured also as acoustically or nurse call system Acoustic if enabled...

Page 13: ...adjusted 9 3 2 Logic nurse call by this button the nurse call logic hardware relais can be permanently turned one or off If the nurse call logic is deactivated the relais will be activated in the case...

Page 14: ...d Time 9 5 Date setup Please see chapter 10 Set up Date and Time 9 6 Touch calibration by using this function the touch screen will be calibrated The operator needs to the crosses in the display withi...

Page 15: ...Page 15 of 23 USER MANUAL ADS intelligent Softwareversion 0 95 Inside this submenu the operator can select one language...

Page 16: ...mode 9 8 1 Bedding mode In this mode the mattress will be inflated with full pressure for maximum 20 minutes After 20 minutes the system turns into the previous mode automatically again 9 8 2 head en...

Page 17: ...s shown in the display and in the schematic mattress is marked by a small red bar above the motion monitor graph 1 By touching the time graph the operator can scroll horizontally from early to late in...

Page 18: ...e current figure which is changed is highlighted in black colour 2 By pressing this button 2 the new settings will be saved 3 By pressing this button the operator will leave this submenu without savin...

Page 19: ...patient monitoring a without redistributing any movement of patient please see chapter Error Reference source not found for details AND b Alternating cycle time mode is enabled One cycle of alternati...

Page 20: ...le format and using the data The data will be saved in CSV format and can be opened at the PC by Excel of Calc One file per day will be saved with the name year month day csv Example a data file with...

Page 21: ...Alarme sind m glich Warnings Priority Pump up 1 The mattress is still into pump up procedure and it is not allowed to place the patient onto the mattress If the pressure inside each cell is sufficient...

Page 22: ...following medical device Product name ADS intelligent Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with risk class 1 bas...

Page 23: ...USER MANUAL ADS intelligent Softwareversion 0 95 manufacturer of ADS intelligent AirMed PLUS GmbH Harpener Feld 34 D 44805 Bochum Germany Tel 0049 0 234 91773320 Fax 0049 0 234 91773322 Email service...

Reviews: