Usage
9.4 Preparation at the operating
location
Wear hand protection�
Wear eye protection�
Use a mask�
Wear protective clothing�
WARNING
Risk of infection from contaminated
products
Risk of cross contamination
i
Reprocess the product correctly and
promptly before its first use and after
every subsequent use�
i
Protect the unit from contamination when
transporting it from the treatment chair to the
reprocessing location�
i
Brush off all surfaces below the water surface
with a soft hygienic brush until they are clean
to the eye�
i
Wipe off all surfaces for at least one minute
with a disinfection wipe�
9.5 Manual cleaning, intermediate
rinsing, disinfection, final rinse,
drying
A combined cleaning agent and disinfectant is
required for manual cleaning and disinfection� It
must have the following properties:
– certified, possibly virucidal efficacy (DVV/RKI,
VAH or European Standards)
– without chlorine, solvents, strong alkaline
solutions (pH >11) or oxidizing agents
For further information, see "8 Cleaning and dis-
infection"�
Cleaning
i
Place individual parts in a cleaning agent bath
making sure that all parts are covered�
i
Comply with the exposure times of the clean-
ing agent�
- Programs:
Cleaning without neutralization
and
THERMAL DES
– Steam sterilization:
- Steam sterilizer Systec DX-45 (Systec
GmbH, Linden, Germany)
9.3 General information
i
Comply with all national directives, standards
and specifications for the cleaning, disinfec-
tion and sterilization of medical devices as well
as the specific specifications for dental prac-
tices and clinics�
i
Comply with the specifications (in "9�5 Manual
cleaning, intermediate rinsing, disinfection, fi-
nal rinse, drying" and "9�6 Automatic cleaning,
intermediate rinsing, disinfection, final rinse,
drying") when you select the cleaning agents
and disinfectants to be used�
i
Comply with the concentrations, tempera-
tures, residence times and post-rinsing speci-
fications issued by the manufacturer of the
cleaning agent and disinfectant�
i
Only use cleaning agents that are non-fixing
and aldehyde-free and display material com-
patibility with the product�
i
Only use disinfectants that are aldehyde-free
and display material compatibility with the
product�
i
Do not use any rinse aid (danger of toxic resi-
due on the components)�
i
Only use freshly-produced solutions�
i
Only use distilled or de-ionized water with a
low bacterial count (at least drinking water
quality) that is free from facultatively patho-
genic microorganisms (e�g� Legionella bacte-
ria)�
i
Use clean, dry, oil- and particle-free com-
pressed air�
i
Do not exceed temperatures of 281 °F�
i
Subject all devices used (e�g� ultrasonic bath,
cleaning and disinfection device (washer-disin-
fector), sealing device, steam sterilizer) to reg-
ular maintenance and inspections�
EN
US
OBJ_DOKU-0000003575-mehrere-005.indd 39
17.04.2019 14:29:04
Summary of Contents for ProVecta 3D Prime
Page 2: ...OBJ_DOKU 0000003575 mehrere 005 indd 2 17 04 2019 14 28 49 ...
Page 4: ...OBJ_DOKU 0000003575 mehrere 005 indd 2 17 04 2019 14 28 49 ...
Page 52: ...OBJ_DOKU 0000003575 mehrere 005 indd 48 17 04 2019 14 29 05 ...
Page 54: ...OBJ_DOKU 0000003575 mehrere 005 indd 2 17 04 2019 14 29 05 ...
Page 104: ...OBJ_DOKU 0000003575 mehrere 005 indd 50 17 04 2019 14 29 11 ...
Page 106: ...OBJ_DOKU 0000003575 mehrere 005 indd 2 17 04 2019 14 29 12 ...
Page 156: ...OBJ_DOKU 0000003575 mehrere 005 indd 50 17 04 2019 14 29 18 ...
Page 157: ...OBJ_DOKU 0000003575 mehrere 005 indd 51 17 04 2019 14 29 18 ...
Page 158: ...OBJ_DOKU 0000003575 mehrere 005 indd 52 17 04 2019 14 29 18 ...
Page 159: ...OBJ_DOKU 0000003575 mehrere 005 indd 53 17 04 2019 14 29 18 ...