TESTING Continued
Turn the “Speed Control” switch clockwise until it clicks and set it to “HIGH”.
The Air Scrubber fan will turn on. If it does not, leave the switch on until you test the
UVC interlock switch. When the UVC lamps are ON, the Blower should run, but
when the HEPA replacement door on the outside is open, the UVC lamps and the
blower will not turn on.