- 15 -
Limit Action opens pop-up menu which defines the action that will take place when the limit
condition is met. This can be an audible beep, a pulse from the trigger socket, or both.
Further actions of Sweep Triggering or Data Logging can be created from the respective
function menus. The action is created at the end of the sweep.
Limit Offset opens a new control screen from which a limit can be offset from its initial
values. This is particularly useful for limit patterns which would otherwise have to be
recreated.
Limit Fix/Unfix
fixes the limit pattern on the screen so that it will not change when the
centre frequency is changed. A message is displayed within the graticule area. This
allows a pattern to be used for alternative\n frequency points. Unfix recalculates the
pattern for the current frequency range and level..
Channel Markers are a special case of a limit pattern file which create only vertical lines at frequency
points defined in the file. Channel marker files are created within PSA Manager and have files names
starting with a $ symbol.
4.6 Traces/Markers
4.6.1 Traces/Markers > Traces Control
Controls the display of the three traces: Live (green), View (white), and Reference
(purple). Also controls the creation of the View trace.
View copies the current Live trace to the View trace, and turns the View trace on if it
was off.
Swap halts the Live trace and places the View trace in front of the Live trace.
Measurement markers then operate on the View trace. Pressing the key again, or
re-starting the Live trace with the Run key will swap the traces back.
View On/Off toggles the View trace on or off without creating
a new View trace. Ref On/Off toggles the Reference trace on
or off. Live On/Off toggles the Live trace on or off. In Dual
Trace mode , both traces are toggled together.
4.6.2 Traces/Markers > Trace Mode
Provides control of the way in which the sweep is written to the Live trace.
Normal each frequency point of the trace is written with the amplitude obtained from
the current sweep.
Peak Hold the trace is written with the highest amplitude obtained from the current
sweep.
Average the trace is written with the highest amplitude obtained from the current
sweeps.
Reset operates in both Average and Peak hold modes to re-start the process from
zero.
More 1 of 2 selects a further set of function keys relating to Dual Trace mode when
either Peak Hold or Average have been selected:
Dual Trace Mode When the trace mode is set to Peak Hold or Average an alternative display
mode can be selected where the normal and processed traces are displayed together.
Limit
Action
Traces
Control
View
Off/On
Trace
Mode
Normal
Reset
Limit
Offset
Peak
Hold
Average
View
Swap
Ref
Off/On
Live
Off/On
More
Limit
Fix/Unfix