aidacare.com.au

7.  Lead the cable towards the remainder of the cables running towards the 

head side of the bed. Zip ties cable to the cables until it reaches the control 

box at the head end of bed. (Only applicable to beds with serial number after 

FL250-V2-2204-24505)

8.  To connect the cord to the control box, remove the control box from the 

mounting bracket by removing the screw holding it in place and sliding 

control box out. (Only applicable to beds with serial number prior  

to FL250-V2-2204-24505)

9.  Open control box and connect the splitter cable to the port shown

NOTE: battery backup (if installed) should be unplugged in order to connect 

the splitter cable. The battery backup is plugged into the splitter cable instead. 

(See next steps)

10. Close control box and reattach to bed
11.  Attach the USB port cable to the splitter cable port marked ‘TFA’ and battery 

backup (if installed) to port marked ‘TBB’

12.  Test bed and USB port functions correctly. Ensure cables have enough slack 

to accommodate all deck panel movements

13.  Ensure cables are secured in place and protected from any moving parts  

of the bed. Zip tie excess cable length to bed frame if required

Your USB upgrade kit installation is now complete.

8. GENERAL OPERATION

8.1 CASTOR SYSTEM – BRAKES & STEERING

8.1.1 Central Locking Castors

Central locking castors are located at the foot end of the bed. The brakes are 

interlinked on either side of the bed allowing operation with a single movement 

for improved safety and efficiency.
1.  To engage the brakes push down on the red pedal as shown below until bed 

is locked in place

2.  To disengage brakes push down on the green pedal until the brake system  

is completely unlocked 

Pedal Positions:

Locked (Red Down)

Unlocked (Green Down)

8.1.2 Directional Locking Castors

Directional locking castors are located at the head end of the bed
1.  To engage the directional lock, ensure the green latch is pushed down  

and secured within the wheel

2.  To disengage the directional lock, pull up on the green latch until unlocked

Locked (Green Down)

Unlocked (Green Up)

Summary of Contents for BEB046100

Page 1: ...FL250 FLOORLINE BED USER MANUAL...

Page 2: ...manual carefully before using your new bed TABLE OF CONTENTS Symbols Used In This Manual 4 Introduction 5 2 Safety Precautions 6 3 Overview 8 4 Specifications 9 5 Accessories 10 6 Electronic Functions...

Page 3: ...angle maximum 35 Bi lateral mattress retainers with length adjustable mattress extension retainer at foot end Integrated foot end extension Extends the bed by 100mm or 200mm Large range of feature en...

Page 4: ...tem must be reset when any changes are made to it Changes are defined as mains power is lost power outage turned off on mains cable is unplugged component cabling is disconnected from or connected to...

Page 5: ...rade IP54 Standards Compliance Aidacare FL250 Beds are designed and manufactured where required in compliance with International Medical Device Safety standard IEC60601 1 IEC60601 2 52 All electronics...

Page 6: ...day chairs Please contact Customer Service on 1300 133 120 for further information Code Description BEA030050 EasyMove Transport System Does not include bed shown BEB046200 BEB046220 Bed End Set Singl...

Page 7: ...button is pressed Adjustment will stop when the button is released or the selection reaches the end of its range Auto Contour Conveniently positions patient into cardiac chair position for therapeutic...

Page 8: ...Adjustable Mattress retainer 1 x Wall Buffer 1 x Handset 1 x Set of 2 Head Board Holders short 2 Foot Board Holders long 1 x Head Board Sold Separately 1 x Foot Board Sold Separately WARNING The FL250...

Page 9: ...cord out as desired d Tighten cable gland to lock cable in place e Test bed through both its hi lo and backrest movement range and ensure the power cord does not get pinched or crushed by the frame An...

Page 10: ...s pre stamped locator points insert the shaft into the bed frame until you feel or hear a click which is the bed frame lining up with the shaft locator point 18 Fasten each Holder at the locator point...

Page 11: ...ch as forklift walkie stacker pallet jack hydraulic lift trolley etc lower the bed carton to 600 700mm from the ground 2 Carefully remove strapping and external dunnage materials remove the carton lid...

Page 12: ...ght 4 Line up the Head Board Bracket Pegs with the Head Board Holder keyholes 5 Insert ensuring both sides are fully inserted push the Head Board down so the pegs drop into the bottom of the Holder ke...

Page 13: ...1 x FOOT section LEFT second longest unique to this section and side 1 x FOOT section RIGHT second longest unique to this section and side Item 2a Hardware for Single Bed 18 x Attachment Bolts and Wa...

Page 14: ...mount the SHP on left or right 5 Undo the left or right Head Board attachment bracket screws and remove the outside plate 6 Position the SHP attachment and receiver bracket over the Head Board holes...

Page 15: ...cers NOTE If Side Panels are already fitted to the single bed they must be removed before the King Single Conversion Kit can be attached Recommended Tools Power Drill Right Angled Power Drill Adaptor...

Page 16: ...nded Tools Power Drill Right Angled Power Drill Adaptor Fastening Folding Assist Bar to Bed 1 For best access to the bed frame side to attached the Folding Assist Bar use the FL250 EasyMove Transport...

Page 17: ...owards the foot end of the bed as shown 4 Remove the 3 x bolts bushes and dome nuts from the mounting plate 5 If mounting to a King Single bed use the bushes supplied For a single bed only the bolts a...

Page 18: ...ocate and unclip the cover exposing existing leads connected to ports and spare port 6 Position the battery back up lead so the groove is aligned with the spare port and insert fully into port 7 Reatt...

Page 19: ...ally so the safety clip is to the right of the quick release axle 9 Lower the bed until the bed frame axle mount is level with the quick release axle on each EasyMove trolley 10 While pressing the qui...

Page 20: ...e 26 If more than one hand is required there is likely to be an encumbrance on the pivot motion please check around the bed 27 When the bed frame is vertical re engage the pivot locks by pulling and t...

Page 21: ...cable to beds with serial number prior to FL250 V2 2204 24505 4 Assemble USB port to the underside of the backrest using the mattress retainer legs as studs and secure in place with handwheels bed ser...

Page 22: ...ort functions correctly Ensure cables have enough slack to accommodate all deck panel movements 13 Ensure cables are secured in place and protected from any moving parts of the bed Zip tie excess cabl...

Page 23: ...s designed for use as a bed extension only and is not meant to support the body weight of a person sitting The extra length of the bed with extra load can also make the bed unstable potentially result...

Page 24: ...leaning NOTE For easy access to all parts of the bed and for safe handling use the EasyMove Transport System NOTE The following procedure also applies to bed accessories but does not include mattresse...

Page 25: ...the control box should always be reset to rule out any calibration issues Calibration issues can occur when power outages happen or when beds are unplugged from the wall during use To complete a reset...

Page 26: ...cts where any original components have been changed or modified by non Aidacare approved and trained service maintenance staff or external providers Furthermore safety is not guaranteed where componen...

Page 27: ...ABUM10 61 281122 AIDACARE PTY LTD Building 3A 1 Moorebank Avenue Moorebank NSW Australia 2170 1300 133 120 aspirecare com au MADE IN PRC...

Reviews: