Section 10
Troubleshooting
Section 10 - Troubleshooting
Basic troubleshooting for BabyPod 2 appears below. If there are questions or concerns not addressed below,
take BabyPod 2 out of service and contact our Customer Service Department.
Problem
Possible cause
Correction
Section
Baby too cold
Disposable warming mattress
not in use.
Activate TransWarmer Mattress before
placing baby in BabyPod 2.
6.3
Baby too cold
Disposable warmer mattress
used in excess of 2 hours.
Activate new TransWarmer Mattress after 2
hours of transport.
6.3
Baby too hot
BabyPod 2 used outside
prescribed environmental
conditions.
DO NOT use the TransWarmer Mattress in
ambient temperatures greater than 28.0
degrees Celsius.
6.3
Visual damage to
BabyPod 2 Outer Shell
Damaged due to sharp
objects or blunt force.
Inspect inside of carbon fibre Outer Shell
for cracks, replace if necessary.
8
Buckle not fully engaged Inadequate connection.
Listen for audible click to engage fully.
6.1
Baby not secure within
patient compartment
Vacuum Mattress not
correctly suctioned.
Suction air from Vacuum Mattress with
baby in place.
6.6
Baby not secure within
patient compartment
Infant Positioning Straps not
properly secured.
Secure Infant Positioning Straps.
6.6
Lid opens during
transport
BabyPod 2 Viewing Lid not
properly secured.
Check that viewing lid securing latch are
properly engaged.
6.7