5. DISPLAY AND KEYBOARD
5.1 Display and Function Selection
The display (5) of the measuring device ALMEMO
®
2290-2/3 consists of an
LCD module with six and a half 7-segment digits, two 16-segment digits, and a
battery symbol and seven arrows for indicating the operating status.
CORR AVERAGE START HOLD RS232C F1 F2
Special Operating Conditions
Segment test of the display automatically after switch-on.
Supply voltage: lower than 7V:
symbol illuminated
BAT
lower than 6V:
1:L o b A t
Sensors that are not connected,
deactivated measuring points,
cleared programming values.
Sensor correction or scaling
arrow
CORR
illuminated.
Measuring point scan in progress arrow
START
illuminated.
Memory for momentary values
arrow
HOLD
illuminated.
Measuring point scan with output
arrow
V24
illuminated.
Measuring function selected
arrow
F1
illuminated.
Programming function selected arrow
F2
illuminated.
Alarm Conditions
are displayed as follows and cause an alarm (see manual 6.3.9):
Sensor breakage:
abbr. flashes
1: N i C r
°C
Exceeding of limit value:
measured value flashes
Overshooting of measuring range:
maximum value flashes
Undershooting of measuring range:
minimum value flashes
Undershoot. of meas.range of CJ comp. or (cold junction)
Measuring without ext. CJC or CJC break.:
flashes
Exceeding of range of values (>65000):
flashes
1:6 5 0 0 0
1: - - - -
XX
1: C J
ALMEMO
2290-2/3
17
Display and Keyboard
Summary of Contents for ALMEMO 2290-2/3
Page 44: ...44 ALMEMO 2290 2 3 Appendix ...