ROUTe Subsystem
10
U2300A Series USB DAQ Programming Guide
95
ROUTe:CHANnel:RVOLtage
Syntax
ROUTe:CHANnel:RVOLtage <value>, (@<ch_list>)
This command is used to set the voltage value of the external reference
ENABlevoltage source of the analog outputs.
ROUTe:CHANnel:RVOLtage? (@<ch_list>)
The query command returns a numeric value that represents the external
reference voltage value in volts for analog outputs.
Supported Model
U2351A, U2353A, U2355A, U2356A, U2331A
Parameters
Remarks
•
It will return an error code if
OUTPut ON
.
•
The reference voltage value cannot be set to zero.
Query Returned Format
<value> [V]
Item
Type
Range of Values
Default Value
<value>
Numeric
From 0 to 10, if the D/A reference voltage source is the
external reference.
10
<ch_list>
Numeric
Single-ended mode:
U2351A/U2352A/U2353A/U2354A: 101-116
U2355A/U2356A/U2331A: 101-164
Differential mode:
U2351A/U2352A/U2353A/U2354A: 101-108
U2355A/U2356A/U2331A: 101-132
N/A
Summary of Contents for U2300A Series
Page 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Page 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Page 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Page 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Page 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Page 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Page 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Page 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Page 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Page 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Page 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...