IEEE-488 Common Command
6
U2300A Series USB DAQ Programming Guide
41
*RCL
Syntax
*RCL {1|2}
This command recalls the instrument state stored in the defined storage
location. It is not possible to recall the instrument state from a storage
location that is empty or was deleted. When shipped from the factory,
storage locations 1 through 2 are empty.
Use the
*SAV
command to store the current instrument state.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Remarks
•
The instrument has two storage locations in the non- volatile memory to
store instrument states. You can only recall a state from a location that
contains a previously stored state.
•
Before recalling a stored state, the instrument will perform the
equivalent of a Factory Reset (
command).
•
A Factory Reset (
command) has no effect on the configurations
stored in the memory. Once a state is stored, it remains until it is
overwritten or specifically deleted.
Example
The following command recalls the instrument state previously stored in
location 1.
*RCL 1
See Also
Summary of Contents for U2300A Series
Page 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Page 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Page 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Page 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Page 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Page 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Page 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Page 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Page 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Page 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Page 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...