TRIGger Subsystem
14
U2300A Series USB DAQ Programming Guide
141
TRIGger:ATRiGger:SOURce
Syntax
TRIGger:ATRiGger:SOURce <src>
This function sets the physical pin for the analog trigger.
TRIGger:ATRiGger:SOURce?
The query command returns a string value representing the trigger source
selection of analog trigger control.
Supported Device
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Remarks
•
Only post and pre- trigger modes can be used if trigger source is set to
SONE
.
•
Input range should be configured well. The
HTHReshold
must be less
than the maximum input range while the
LTHReshold
must be larger
than minimum input range.
•
To activate
SONE
trigger, users are required to run the
command
before or after analog output (
"OUTPut ON"
).
•
The
SONE
trigger detection only comes into effect during the
command execution.
•
The
SONE
parameter refers to the first entry in
command.
Query Returned Format
<src> ::= {EXTAP|SONE}
Item
Type
Range of Values
Default Value
<src>
String
EXTAP: From external analog trigger pin
SONE: Source from first scanned channel for
multiplexing DAQ device
EXTAP
Summary of Contents for U2300A Series
Page 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Page 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Page 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Page 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Page 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Page 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Page 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Page 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Page 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Page 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Page 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...