17 Technical Information
Appendix 2
394/400
TPS-mobile User Manual / 87-901-005-01 (F)
Appendix 2
Symptom
Possible cause
Action
System doesn’t start
Problems on main connection
Check the connection of main
plug.
Check possible fuses on main
line
Main switch placed on position “0”
Move the Main Switch on
position “1”
Turbo pump is not able to reach
the “Normal Operation Status”
Leak on vacuum system connected
to the inlet flange
Check the vacuum system
critical points (valves, flanges,
plastic pipes,…)
TMP Vent Valve not properly closed
Check if the vent valve is
screwed completely down
Leak on Vent Valve
Check the Vent valve sealing o-
ring positioning and intactness
TMP stopped and the Controller
shows the message “Pump
Overtemp”
Room temperature higher than 35 °C [95 F]
Summary of Contents for TPS-mobile 9698403
Page 3: ...TPS mobile TPS mobile User Manual 87 901 005 01 F 3 400 TPS mobile...
Page 4: ...TPS mobile 4 400 TPS mobile User Manual 87 901 005 01 F...
Page 34: ...1 Istruzioni per l uso Smaltimento 34 400 TPS mobile User Manual 87 901 005 01 F...
Page 56: ...2 Gebrauchsanleitung Entsorgung 56 400 TPS mobile User Manual 87 901 005 01 F...
Page 156: ...7 Istruktionsbog Bortskaffelse 156 400 TPS mobile User Manual 87 901 005 01 F...
Page 176: ...8 Bruksanvisning Bortskaffning 176 400 TPS mobile User Manual 87 901 005 01 F...
Page 196: ...9 Instruksjon Manual Kassering 196 400 TPS mobile User Manual 87 901 005 01 F...
Page 216: ...10 Ohjek sikirja H vitt minen 216 400 TPS mobile User Manual 87 901 005 01 F...
Page 276: ...13 N vod k Pou it Likvidace 276 400 TPS mobile User Manual 87 901 005 01 F...
Page 296: ...14 N vod na Obsluhu Likvid cia 296 400 TPS mobile User Manual 87 901 005 01 F...
Page 395: ......
Page 396: ......
Page 397: ......
Page 398: ......
Page 399: ......