Step 1.
Press
Mode
,
TD-SCDMA with HSPA/8PSK
to enable the TD-SCDMA mode
measurements.
Step 2.
Press
Mode Preset
to preset the analyzer.
Step 3.
Press
Input/Output
,
External Gain
,
BTS
and enter the number and
−
dB
according
to the external attenuator connected to the analyzer.
Step 4.
Press
Mode Setup
,
Radio
,
Device
to toggle the device to
BTS
.
Step 5.
Press
Meas
,
Monitor Spectrum
to initiate the Monitor Spectrum measurement.
Step 6.
Press
Sweep/Control
,
Gate
and toggle the
Gate
to
On
. Press
More
,
Gate Source
and select the desired gate source
External 1
.
If no measurement is performed and only a symbol of yellow ‘*’ displays on the
top right of the screen, you need to check the gate settings. You need the correct
gate signal to make the Monitor Spectrum measurement.
Step 7.
Press
FREQ Channel
,
2
,
GHz
to set the center frequency to 2 GHz.
Step 8.
Press
SPAN X Scale
, enter a numerical span using the front-panel keypad, and
select a units key, such as
MHz
.
NOTE
A display with a Spectrum window appears when you activate a Spectrum
measurement. Changes to the
FREQ
,
Span
, or
AMPTD
settings will affect only the
active window.
Step 9.
Press
Meas Setup
to check the keys available to change the measurement
parameters from the default condition.
Step 10.
Press
Trace/Detector, Select Trace
and select the trace(s) desired for display, then
toggle
Display
to
Show
.
For more details about changing measurement parameters, see
N9079A
TD-SCDMA with HSPA/8PSK User’s and Programmer’s Reference
.
If you have a problem, and get an error message, see
the “
Instrument Messages
Guide
”
.
Summary of Contents for N9010A
Page 4: ......
Page 7: ...Contents 7 Waveform Signal Envelope Views of I only or Q only 106...
Page 8: ...8 Contents...
Page 12: ...12 Chapter1 Making TD SCDMA with HSPA 8PSK Measurements Setting Up and Making a Measurement...
Page 25: ...Chapter 3 25 ACP Measurements Measurement Procedure Messages Guide...
Page 28: ...28 Chapter3 ACP Measurements Measurement Results...
Page 44: ...44 Chapter5 Spectrum Emission Mask Measurements Troubleshooting Hints...
Page 64: ...64 Chapter8 Power Statistics CCDF Measurements Troubleshooting Hints...
Page 94: ...94 Chapter12 Monitor Spectrum Measurements Measurement Results...