6
Site Preparation
Introduction
This section discusses the site preparation for the HP 5890 SERIES II
(hereafter referred to as HP 5890). The following topics are discussed:
C
Environmental considerations
C
Power considerations
C
Gas considerations
You will need various supplies to prepare an area for the HP 5890.
Hewlett•Packard's analytical supplies catalog describes the supplies
available from Hewlett•Packard.You can obtain a copy from your local
sales office.
Environmental considerations
This instrument operates in the 0
^
to 55
^
C temperature range and the
5% to 95% relative humidity range. We recommend an environment
comfortable for the operators (reasonably constant temperature and
humidity) for optimum performance and instrument lifetime.
Corrosive substances, whether gas, liquid, or solid, may affect materials
and components used in the HP 5890 and should be avoided.
Space considerations
The HP 5890 requires a clear surface about 56 cm (22 in.) deep by 71 cm
(28 in.) wide, capable of supporting at least 41 kg (90 lbs). Another 25 cm
(10 in.) must be allowed behind the instrument to dissipate oven exhaust
air (up to 400
^
C).
Summary of Contents for HP 5890 II Plus Series
Page 1: ...Chemical Analysis Group 5890 Gas Chromatograph 5890 Series II Site Prep Info Document A15282 ...
Page 2: ...Site Prep Installation Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 6: ...1 Site Preparation ...
Page 19: ...This page intentionally left blank ...
Page 20: ...2 Installation ...