Installation
Connecting signal output cables
40
Note that the
general purpose
and
HP 3350Series LAS
analog output
cable assemblies consist of two independent cables, terminating together
at a common, single,
female
plug at the HP 5890. One cable is labelled
1
mV
, the other
+1 V
output. In general, the
+1 V
cable is connected to an
integrating device or A/D converter, and the
1 mV
cable is connected to a
chart recording device.
Part No.
Typical Use
Length
Connector Type
35900-60610
to HP 3394/96A
2 m
Edge Connector:
Integrator
HP 3394/96A Integrator,
1 V input
05890-60760
to HP 3390/92/93A
2 m
Edge Connector:
Integrator
HP 3390/92A Integrator,
1 V input
05890-60790
to HP 3350 Series
2 m
Edge Connector:
Lab Automation
HP 18652A A/D Converter,
Systems
1 V input
05890-60780
to HP 3388A
2 m
Push-On Terminals:
Integrator
1 V A/D inputs
05890-60800
General Purpose
2 m
Spade Lugs
Note: Information above pertains to the +1 V cable. The 1 mV cable terminates
in spade lugs for cable Part No. 05890-60790 and -60800.
Analog Signal Output Cables
1. Insert the cable plug into the
SIGNAL 1
or
SIGNAL 2
male
receptacle
.
The plug may be inserted in either of its two possible
orientations.
Note:
The
SIGNAL 2
analog output receptacle does not function unless
the Option 550/Accessory 19242A (Communications Interface Board) or
Option 560/Accessory 19254A (RS•232•CInterface Board) is installed.
Summary of Contents for HP 5890 II Plus Series
Page 1: ...Chemical Analysis Group 5890 Gas Chromatograph 5890 Series II Site Prep Info Document A15282 ...
Page 2: ...Site Prep Installation Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 6: ...1 Site Preparation ...
Page 19: ...This page intentionally left blank ...
Page 20: ...2 Installation ...