Installation
Connecting supply gases
29
Assemble fitting,
bottom tubing
Hand-tighten, then
wrench-tighten
C
For a carrier gas
, its tube is connected at the fitting on the
appropriate mass flow controller labeled
IN
.
C
For a capillary makeup gas
, its tube is connected at the fitting on
the appropriate flow manifold block labeled
FID—AUX
for an FID or
NPD, or
TCD—AUX
for a TCD or ECD.
C
For H
1
and air for an FID or NPD
, their tubes are connected at
fittings on the appropriate flow manifold block labeled, respectively,
FID—HYD
and
FID—AIR
.
C
For TCD reference gas
, its tube is connected at the fitting on the
appropriate flow manifold block labeled
TCD—REF
.
Once all connections are made, proceed to Setting initial supply
pressures and leak testing.