Getting Started
Agilent Electrothermal Temperature Controller ETC 60 User’s Guide
22
9.
Screw the thumbscrew through the bracket support and into the side of the
burner adjuster to hold the bracket in place.
WARNING
Sharp Objects
Broken glass can cause severe cuts to hands. Handle the workhead with
care, as the sidearm of the quartz tube is fragile and easily broken.
10.
Install the heat shield and secure it with the two screws.
11.
Go to ‘Installing the Workhead’ on Page 23.
ETC 60 Workhead Bracket Installation on Zeeman Instruments
Additional steps must be performed before installing the ETC workhead bracket on
a Zeeman instrument.
1.
Remove the PSD 120 autosampler (if present)
2.
Remove the Zeeman workhead from the instrument.
Refer to the 240/280 instrument user manuals for detailed instructions.
Cell Holder and Adapter Removal
The standard cell holder must be removed to allow installation of the ETC 60
Workhead.
Figure 21.
Adapter, standard cell holder and cell inside sample compartment
1.
Lift the cell holder up to unhook cell holder from the adapter mounting.
2.
Lift the adapter out of the sample compartment carefully avoiding the crossbar.
Adapter
Cell
Standard Cell Holder