Agilent E8663B Analog Signal Generator User’s Guide
69
Troubleshooting
No Modulation at the RF Output
4.
Turn the RF off: set
RF On/Off
to On.
5.
Turn the signal generator’s automatic leveling control (ALC) off: press
Amplitude
>
ALC Off On
to Off.
6.
Monitor the RF output amplitude as measured by the power meter.
7.
Press
Amplitude
and adjust the signal generator’s RF output amplitude until the desired power is
measured by the power meter.
Setting Power Search Mode
Power search mode executes a power search routine that temporarily activates the ALC, calibrates the
power of the current RF output, and then disconnects the ALC circuitry. See the
E8663B Analog
Signal Generator Key Reference
for more information on the Power Search function.
Use the following steps to set the signal generator to manual fixed power search mode:
1.
Preset the signal generator: press
Preset
.
2.
Set the desired frequency: press
Frequency
and enter the desired frequency.
3.
Set the desired amplitude: press
Amplitude
and enter the desired amplitude.
4.
Turn the signal generator’s automatic leveling control (ALC) off: press
Amplitude
>
ALC Off On
to Off.
5.
Turn the RF on: set
RF On/Off
to On.
6.
Press
Do Power Search
.
This executes the manual fixed power search routine, which is the default mode.
There are three power search modes: manual, automatic, and span.
When
Power Search
is set to Manual, pressing
Do Power Search
executes the power search calibration
routine for the current RF frequency and amplitude. In this mode, if there is a change in RF
frequency or amplitude, you will need to press
Do Power Search
again.
When
Power Search
is set to Auto, the calibration routine is executed whenever the frequency or
amplitude of the RF output is changed.
When
Power Search
is set to Span, pressing
Do Power Search
executes the power search calibration routine
over a selected range of frequencies at one time. The power search corrections are then stored and
used whenever the signal generator is tuned within the selected range of frequencies.
No Modulation at the RF Output
Check the
MOD ON/OFF
annunciator on the display. If it reads
MOD OFF
, press
Mod On/Off
to toggle the
modulation on.
Although you can set up and enable various modulations, the RF carrier is modulated only when you
have also set
Mod On/Off
to On.