30
Agilent E8663B Analog Signal Generator User’s Guide
Basic Operation
Configuring the RF Output
7.
Press
Edit Item
>
545
>
MHz
.
This changes the frequency value in row 4 to 545 MHz.
8.
Highlight any column in the point 7 row and press
Insert Row
.
This adds a new point between points 7 and 8. A copy of the point 7 row is placed between
points 7 and 8, creating a new point 8, and renumbering the successive points.
9.
Highlight the frequency item for point 8, then press
Insert Item
.
Pressing
Insert Item
shifts frequency values down one row, beginning at point 8. Note that the original
frequency values for both points 8 and 9 shift down one row, creating an entry for point 10 that
contains only a frequency value (the power and dwell time items do not shift down).
The frequency for point 8 is still active.
10. Press
590
>
MHz
.
11. Press
Insert Item
>
-2.5
>
dBm
.
This inserts a new power value at point 8 and shifts down the original power values for points 8 and 9
by one row.
12. Highlight the dwell time for point 9, then press
Insert Item
.
A duplicate of the highlighted dwell time is inserted for point 9, shifting the existing value down to
complete the entry for point 10.
To Configure a Single List Sweep
1.
Press
Return
>
Sweep
>
Freq & Ampl
This turns the sweep on again. No errors should occur if all parameters for every point have been
defined in the previous editing process.
2.
Press
Single Sweep
.
The signal generator will single sweep the points in your list. The
SWEEP
annunciator activates during
the sweep.
3.
Press
More (1 of 2)
>
Sweep Trigger
>
Trigger Key
.
This sets the sweep trigger to occur when you press the
Trigger
hardkey.
4.
Press
More (2 of 2)
>
Single Sweep
.
This arms the sweep. The
ARMED
annunciator is activated.
5.
Press the
Trigger
hardkey.
The signal generator will single sweep the points in your list and the
SWEEP
annunciator will be
activated during the sweep.
To Configure a Continuous List Sweep
Press
Sweep Repeat Single Cont
.
This toggles the sweep from single to continuous. A continuous repetition of the frequencies and
amplitudes configured in the list sweep are now available at the RF OUTPUT connector. The
SWEEP
annunciator appears on the display, indicating that the signal generator is sweeping and progression of the
sweep is shown by a progress bar.