Chapter 5
333
Language Reference
SENSe Subsystem
Spectrum—FFT Minimum Points in Resolution BW
[:SENSe]:SPECtrum:FFT:RBWPoints <real>
[:SENSe]:SPECtum:FFT:RBWPoints?
Set the minimum number of data points that will be used inside the
resolution bandwidth. The value is ignored if length control is set to
manual. This is an advanced control that normally does not need to be
changed.
Factory Preset
and *RST:
1.30
Range:
0.1 to 100
Remarks:
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Spectrum—Window Delay
[:SENSe]:SPECtrum:FFT:WINDow:DELay <real>
[:SENSe]:SPECtrum:FFT:WINDow:DELay?
Set the FFT window delay to move the FFT window from its nominal
position of being centered within the time capture. This function is not
available from the front panel. It is an advanced control that normally
does not need to be changed.
Factory Preset
and *RST:
0
Range:
−
10.0 to +10.0s
Default Unit:
seconds
Remarks:
To use this command, the Service mode must be
selected with INSTrument:SELect. In Service mode, it
is possible to get an acquisition time that is longer than
the window time so that this function can be used.
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...