262
Chapter 5
Language Reference
SENSe Subsystem
SENSe Subsystem
Sets the instrument state parameters so that you can measure the
input signal.
Adjacent Channel Power Measurement
Commands for querying the adjacent channel power measurement
results and for setting to the default values are found in the
Group of Commands” on page 233
. The equivalent front panel keys for
the parameters described in the following commands, are found under
the
Meas Setup
key, after the
ACP
or
ACPR
measurement has been
selected from the
MEASURE
key menu.
Adjacent Channel Power—Average Count
[:SENSe]:ACP:AVERage:COUNt <integer>
[:SENSe]:ACP:AVERage:COUNt?
Set the number of data acquisitions that will be averaged. After the
specified number of average counts, the average mode (termination
control) setting determines the average action.
Factory Preset
and *RST:
10 for cdma2000, W-CDMA (3GPP), W-CDMA (Trial &
Arib)
20 for Basic, cdmaOne, iDEN
Range:
1 to 10,000
Remarks:
Use INSTrument:SELect to set the mode.
Adjacent Channel Power—Averaging State
[:SENSe]:ACP:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:ACP:AVERage[:STATe]?
Turn average on or off.
Factory Preset
and *RST:
On
Off for iDEN mode
Remarks:
Use INSTrument:SELect to set the mode.
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...