Chapter 5
191
Language Reference
CALCulate Subsystem
Marker On/Off
:CALCulate:<measurement>:MARKer[1]|2|3|4[:STATe] OFF|ON|0|1
:CALCulate:<measurement>:MARKer[1]|2|3|4[:STATe]?
Turns the selected marker on or off.
The marker must have already been assigned to a trace. Use
:CALCulate:<measurement>:MARKer[1]|2|3|4:TRACe
to assign a
marker to a particular trace.
Example:
CALC:SPEC:MARK2: on
Remarks:
The keyword for the current measurement must be
specified in the command. (Some examples include:
SPECtrum, AREFerence, WAVeform)
The WAVeform measurement only has two markers
available.
Front Panel
Access:
M
arker, Select
then
Marker Normal
or
Marker On Off
Marker to Trace
:CALCulate:<measurement>:MARKer[1]|2|3|4:TRACe <trace_name>
:CALCulate:<measurement>:MARKer[1]|2|3|4:TRACe?
Assigns the specified marker to the designated trace. Not all types of
measurement data can have markers assigned to them.
Example:
With the WAVeform measurement selected, a valid
command is
CALC:SPEC:MARK2:TRACE rfenvelope
.
Range:
The names of valid traces are dependent upon the
selected measurement. See the following table for the
available trace names. The trace name assignment is
independent of the marker number.
Remarks:
The keyword for the current measurement must be
specified in the command. (Some examples include:
SPECtrum, WAVeform)
Front Panel
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...