
Using the Oscilloscope
48
To use the quick measurements
The action taken when the
[Multi Purpose]
key is pressed depends on the feature
selected in the Customize Multipurpose dialog box. The default feature is
QuickMeas (quick measurements), which is described as follows.
• To turn on the quick measurement display, press the
[Multi Purpose]
key.
The five preset measurements defined in the Quick Measurement
configuration are enabled and results are displayed on the screen for the first
waveform source. The default measurements are: V
p-p
, Period, Frequency, Rise
time, and Fall time.
• To measure parameters for another waveform, press the
[Multi Purpose]
key
until that waveform is the one shown in the measurement readout.
Continuing to press the
[Multi Purpose]
key cycles through each of the
waveforms available.
• To turn off the quick measurement display, press and release the
[Multi Purpose]
key until the measurements are turned off.
The measurement results disappear from the screen.
See the Infiniium application’s online help (described in Chapter 4) for
information on how to configure the quick measurement capability, using the
Customize Measurement feature of the graphical interface.
Summary of Contents for DSO9024H
Page 7: ...1 Setting Up the Oscilloscope...
Page 27: ...2 Working in Comfort...
Page 31: ...3 Using the Oscilloscope...
Page 88: ...Using the Oscilloscope 88...
Page 89: ...4 Accessing the Online Help...
Page 100: ...Index 100...