Setting Up the Oscilloscope
25
Changing Windows Operating System Settings
Many Windows operating system settings can be changed to suit your own
personal preferences. However, some operating system settings should not be
changed because doing so would interfere with the proper operation of the
oscilloscope.
• Do not change the Power Options.
• Do not change the System Properties Hardware Tab settings.
• Do not change the Regional and Language Options Advanced Tab settings.
• Do not remove Fonts.
• Display Settings:
• Do not change the screen resolution or the color quality from Highest (32
bit).
• Do not change the Font size to Extra Large.
• Do not use a Menu font size greater than 14 points.
• Do not use the Administrative Tools to enable or disable Internet Information
Services (Web Server). Use the Infiniium Web Control dialog box to enable or
disable the Web Server.
• Do not delete or modify the Infiniium Administrator user account.
Before changing any Windows operating system settings outside of the
oscilloscope application you should exit the oscilloscope application.