143
Compatible SCPI Commands
Supported SCPI Commands
adjacent-channel power measurements.
[SENSe<1|2>:]POWer:ACHannel:ACPairs 0 | 1 | 2 | 3
Specifies the number of adjacent channels (upper and lower channel in pairs).The number 0 represents a
pure channel power measurement.
[SENSe<1|2>:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel] 100 Hz to 1000 MHz
Specifies the radio communication system’s channel bandwidth.
Remote Command
[:SENSe]:POWer:ACHannel:TXCHannel:COUNt <int>
[:SENSe]:POWer:ACHannel:TXCHannel:COUNt?
Example
POW:ACH:TXCH:COUN 1
POW:ACH:TXCH:COUN?
Remote Command Notes
FSP, FSU
Preset
4
State Saved
Saved in instrument state.
Min
1
Max
4
Remote Command
[:SENSe]:POWer:ACHannel:ACPairs
[:SENSe]:POWer:ACHannel:ACPairs?
Example
POW:ACH:ACP 3
Remote Command Notes
FSE, FSP, FSU
Preset
0
Range
0|1|2|3
Remote Command
[:SENSe]:POWer:ACHannel:BANDwidth|BWIDth[:CHANnel]
<freq>
[:SENSe]:POWer:ACHannel:BANDwidth|BWIDth[:CHANnel]?
Example
POW:ACH:BWID 30kHz
Remote Command Notes
FSE, FSP, FSU
Preset
14KHz
State Saved
Saved in instrument state.
Min
100Hz
Max
1000MHz
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html