69
Compatible SCPI Commands
Supported SCPI Commands
The query returns the current state of this setting.
CALCulate<1|2>:LIMit<1...6>:FAIL?
Returns the limit check result for the specified limit line.
To obtain a valid result, you must perform a complete sweep.
The numeric value returned has the following meaning:
CALCulate<1|2>:LIMit<1...6>:CLEar[:IMMediate]
Deletes the current limit check result for all limit lines.
CALCulate<1|2>:LIMit<1...6>:COMMent <string>
The command defines a comment for the specified limit line. The comment string may have a maximum
length of 40 characters.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:UNIT
DBM|DBUV|DBMV|DBUA|A|V|W
:CALCulate[1]:LIMit[1]|2|3|4|5|6:UNIT?
Example
CALC:LIM:UNIT DBM
CALC:LIM:UNIT?
Remote Command Notes
FSE, FSP, FSU
Preset
DBM
State Saved
Saved in instrument state.
Range
DBM | DBUV | DBMV | DBUA | A | V | W
0
PASS
1
FAIL
2
MARGIN
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:FAIL?
Example
CALC:LIM:FAIL?
Remote Command Notes
FSE, FSP, FSU
Min
0
Max
2
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CLEar[:IMMediate]
Example
CALC:LIM:CLE
Remote Command Notes
FSE, FSP, FSU
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html