162
Compatible SCPI Commands
Supported SCPI Commands
TRIGger Subsystem
This subsystem includes commands that control and synchronize the start of a sweep.
An external trigger signal can be supplied via the connector on the analyzer’s rear panel.
TRIGger<1|2>[:SEQuence]:SOURce
Selects the trigger source.
This mode supports only parameter options: IMMediate, EXTernal, VIDeo, RFPower.
It does
not
support parameter options: IFPower, TV, AF, FM, AM and PM.
TRIGger<1|2>[:SEQuence]:LEVel:IFPower –30 to –10DBM
This command sets the level of the IF power trigger source.
This command is accepted, but takes no action and reports no error.
TRIGger<1|2>[:SEQuence]:LEVel:RFPower
Sets the RF power trigger level.
Preset
OFF
State Saved
Saved in instrument state.
Remote Command
:TRIGger[1][:SEQuence]:SOURce
IMMediate|EXTernal|VIDeo|IFPower|RFPower|TV|AF|FM|AM|PM
:TRIGger[1][:SEQuence]:SOURce?
Example
TRIG:SOUR EXT
Remote Command Notes
Remote Command Notes
FSE, FSP, FSU
Preset
IMMediate
Range
IMMediate|EXTernal|VIDeo|IFPower|RFPower|TV|AF|FM|AM|PM
Remote Command
:TRIGger[1][:SEQuence]:LEVel:IFPower <ampl>
:TRIGger[1][:SEQuence]:LEVel:IFPower?
Example
TRIG:LEV:IFP –20DBM
Remote Command Notes
FSP, FSU
Preset
–20dBm
State Saved
Saved in instrument state.
Min
30dBm
Max
–70dBm
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html