Maintenance and Spare Parts
Cary Eclipse Fluorescence Spectrophotometer User’s Guide
39
Checking
To check a fuse:
1
Disconnect the Cary Eclipse instrument from the mains power
supply.
2
Open the hinged cap and pull out the fuse holder (lower red
section). Both fuses are held in this fuse holder.
3
Check that each fuse is the correct type and that they are not
damaged. If necessary, replace the fuse (refer to Page 38).
4
Place the fuses into the red fuse holder and push this red fuse
holder into the main fuse holder and then lower the cap.
5
Reconnect the Cary Eclipse instrument to the mains power
supply and turn on the instrument.
NOTE
If a fuse repeatedly blows, it may indicate other problems with the Cary Eclipse
instrument.
Spare Parts
Please see the Agilent website for a complete listing of spare parts:
www.agilent.com