Introduction
18
Cary Eclipse Fluorescence Spectrophotometer User’s Guide
Conventions
The following conventions have been used in procedures throughout
the documentation:
Menus, menu items, buttons and check boxes have been typed in
bold
. (for example, ‘click
OK’
and ‘From the
Edit
menu, choose
Copy
’.
ALL CAPITALS indicate keyboard commands. For example,
‘press ENTER’ and ‘press SHIFT+F8’.
Specifications
The instrument is suitable for indoor use
only
and is classified
Pollution degree 2 and Installation Category II (EN 61010-1).
Environmental
Table 1.
Suitable
conditions during instrument transportation, non-operation and operation
Condition
Altitude (m, ft)
Temperature
(°C, °F)
Relative humidity,
non-condensing (%)
Non-operating (transport)
0–2133, 0–7000
5–45, 41–113
20–80
Operating within performance
specifications
0–853, 0–2800
853–2133, 2800–7000
10–35, 50–95
10–25, 50–77
8–80
For optimum analytical performance, it is recommended that the
ambient temperature of the work area be between 20 and 25 °C (68
and 77 °F) and be held constant to within ±2 °C throughout the
entire working day.