Fuses
The Cary Eclipse spectrophotometer contains two fuses that are
located at the back of the instrument.
Replacing
To replace a fuse:
1
Disconnect the Cary Eclipse instrument from the power supply.
2
Replace the blown fuse with one of the same type and rating, as
indicated in Chapter 2.
The fuses have a code marked on the cap (for example, T 3.15 AH
250 V). This refers to the fuse characteristic (‘T’: time lag, ‘F’: fast-
acting), the current rating (‘x’: amperes), the braking capacity (‘H’:
heavy, ‘L’: low) and the voltage rating (‘y’: volts).
NOTE
Fuse information on the rear of the instrument is the most up-to-date.
WARNING
Electrical Shock and Fire Hazards
To prevent reduced safety protection or unwanted fusing,
always
ensure
that the code on the fuse cap matches the information printed next to the
fuse holders.