![Agilent Technologies Agilent 86120B User Manual Download Page 228](http://html.mh-extra.com/html/agilent-technologies/agilent-86120b/agilent-86120b_user-manual_2867999228.webp)
6-10
Performance Tests
Test 5. Amplitude Accuracy and Linearity
After completing this step, the first two columns of the table should be com-
pletely filled in.
10
Disconnect the fiber-optic cable from the optical power meter and connect it
to the Agilent 86120B’s
OPTICAL INPUT
connector.
11
Set the optical attenuator for the value that you recorded in
Step 8
.
12
Place the polarization controller in the auto scan mode.
13
Press the Agilent 86120B’s front-panel
Preset
key.
14
Press
List by Power
,
Appl’s,
and then
DRIFT
.
15
After two minutes, stop the polarization controller’s auto scan function.
16
Press the
MAX-MIN
softkey so that
MAX
is highlighted. Enter the maximum drift
reading on the following line:
maximum drift: ____________
17
Press the
MAX-MIN
softkey so that
MIN
is highlighted. Enter the minimum drift
reading on the following line:
minimum drift: ____________
18
, and
to calculate the power-
correction offset value as shown in the following equation:
Enter the calculate value on the following line:
power-correction offset: ____________
19
Change the attenuator to the settings shown in
. For each setting,
record the power measured on the Agilent 86120B.
After completing this step, the table’s column titled “Agilent 86120B Power
Reading” should be completely filled in.
20
Calculate the “Linearity” value for each row in the table using the following
equation:
21
Compare the linearity values with the specification listed in
“Specifications and Regulatory Information”
. The data may show multiple
offset
minimum drift
maximum drift
+
2
-------------------------------------------------------------------------------------------
Pwr
–
=
Linearity
Power Meter Reading
HP 86120B Power Reading
–
offset
–
=
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......