5-42
Programming Commands
CALCulate2 Subsystem
WLIMit:STOP:WNUMber
Sets the stopping wavenumber for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP:WNUMber {?|{<real> | MINimum | MAXimum }}
<
real
> is a wavenumber value that is within the following limits:
Constant
Description
MINimum start
wavelength
limit
MAXimum 14286
cm
-1
(700 nm)
Attribute
Summary
Non-sequential command
Preset
State: 8.333335E5 m
-1
*RST State: 8.333335E5 m
-1
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The default
units for the <
real
> parameter are inverse meters. The stop wavenumber
value must be less than or equal to the start wavenumber value or the stop
wavenumber will be clipped to the start wavenumber and a “Data out of
range” error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......