5-71
Programming Commands
SENSe Subsystem
CORRection:ELEVation
Sets the elevation value used by the instrument to compensate for air disper-
sion.
Syntax
:SENSe:CORRection:ELEVation{?| {<integer> | MINimum | MAXimum}}
<integer>
is the altitude in meters.
Constant
Description
MINimum 0
m
MAXimum 5000
m
Attribute
Summary
Non-sequential command
Preset
State: unaffected by
*RST sets this value to the minimum.
SCPI Compliance: instrument specific
Description
Changing the elevation value causes the current data to be reprocessed.
Query Response
The query form returns the current elevation setting as shown in the following
example:
+1500
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.