![Agilent Technologies Agilent 86120B User Manual Download Page 180](http://html.mh-extra.com/html/agilent-technologies/agilent-86120b/agilent-86120b_user-manual_2867999180.webp)
5-48
Programming Commands
CALCulate3 Subsystem
DELTa:REFerence:FREQuency
Selects the reference laser line for DELTa calculations.
Syntax
:CALCulate3:DELTa:REFerence:FREQuency{?| {<real> | MINimum | MAXimum}}
<real>
is a frequency value that is within the following limits:
Constant
Description
MINimum 181.6924
THz
MAXimum 428.6
THz
Attribute
Summary
Preset
State: 428.6 THz (700 nm)
*RST State:428.6 THz (700 nm)
SCPI Compliance: instrument specific
Description
The reference will be the laser line at the frequency closest to the frequency
entered. Subsequent measurements will use the frequency closest to the ref-
erence frequency used for the previous measurement.
The query returns the reference laser line’s frequency. The default units for
the
<real>
parameter are Hz.
DELTa:REFerence:POWer?
Queries the reference laser line’s power level.
Syntax
:CALCulate3:DELTa:REFerence:POWer?
Attribute
Summary
Preset
State: not affected
*RST State: not affected
SCPI Compliance: instrument specific
Query Only
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......