![Agilent Technologies AdvanceBio Amino Acid Analysis User Manual Download Page 5](http://html1.mh-extra.com/html/agilent-technologies/advancebio-amino-acid-analysis/advancebio-amino-acid-analysis_user-manual_2867682005.webp)
Other Operating Tips
(continued)
• Columns should not be maintained at elevated pH or elevated
temperature when not in use (see
Table 2
).
• New columns may contain a mixture of organic solvents and water,
which may contain buffer salts. See
Table 3
for details of the shipping
solvents. Initially, care should be taken not to pass any mobile phase
through the column that may cause a precipitate to form or may not
be fully miscible.
• Agilent BioHPLC and AdvanceBio columns are compatible with all
common eluents used for analysis of biomolecules.
ENGLISH
Column
Particle size
Pressure limit
AdvanceBio Peptide Mapping
2.7 µm
600 bar
AdvanceBio Peptide Plus
AdvanceBio Oligonucleotide
AdvanceBio Amino Acid Analysis
AdvanceBio RP-mAb
3.5 µm
ZORBAX 300SB
3.5 µm, 5 µm, 7 µm
400 bar
Poroshell 300
5 µm
ZORBAX RRHD
1.8 µm
1200 bar
PLRP-S
3 µm
275 bar
5 µm, 8 µm, 10 µm
207 bar
10-15 µm, 15-20 µm, 30 µm
103 bar
Table 1. Maximum operating parameters – columns
up to 4.6 mm
Summary of Contents for AdvanceBio Amino Acid Analysis
Page 22: ...1 pH 2...
Page 27: ...20 30 50 50 20 30 5 75 25 6 100 100 100 PLRP S 80 1 M NaOH 1 M HCI 20 1 8 m...
Page 56: ...Agilent BioHPLC AdvanceBio www agilent com chem jp QC Agilent QC LC QC www agilent com chem jp...
Page 58: ...1 pH 2...
Page 62: ...1 8 m 0 5 m AdvanceBio 2 m 2 m www agilent com chem jp 1 8 m 1 2 10 20 3 100 4...
Page 63: ...20 30 50 50 20 30 5 75 25 6 100 100 100 PLRP S 1 M 1 M 80 20 1 8 m...
Page 67: ...Agilent 1 pH 2...
Page 71: ...1 8 0 5 AdvanceBio Peptide Mapping 2 2 www agilent com chem guards 1 8 1 2 10 20 3 100 4...
Page 72: ...50 50 20 30 20 30 5 75 25 6 100 100 100 PLRP S 80 1 M NaOH 1 M HCI 20 1 8...