Technical Information
Turbo-V 1K-G Navigator Controller Installation
17
Turbo-V 1K-G User Manual
/
87-901-025-01 (A)
293/304
Fore-Vacuum Pump Connection
A flange KF 40 NW (ISO 200 F inlet flange) and KF 25 NW (ISO 160 F
inlet flange) are available to connect the Turbo-V 1K-G pump to the
fore-vacuum pump. A hose or vacuum approved pipe can be used. If
a rigid pipe is used, any vibration generated by the mechanical pump
must be eliminated through the use of bellows.
NOTE
The Turbo-V 1K-G pump is characterized by its high compression ratio also for
oil vapors. When using a mechanical oil-sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore-vacuum pump in order to
prevent oil backstreaming.